80PK-8 Fluke, 80PK-8 Datasheet

Test Probe

80PK-8

Manufacturer Part Number
80PK-8
Description
Test Probe
Manufacturer
Fluke
Type
Pipe Clamp Prober
Datasheet

Specifications of 80PK-8

Test Probe Functions
Temperature
Range
-20deg To +300degF
Accuracy
± 1.9°C (± 3.4°F)
Color
Black
Conductor Size
24 AWG
Connector Type
Mini-Thermocouple
Length, Cable
1 m
Material, Handle
PBT
Measuring Range
-29 to 149 –29 to 149°C
Plug Type
0.792 mm Mini Thermocouple
Used With
Infrared Themometer
Voltage, Range, Ac
24 V
Voltage, Range, Dc
60 V
Voltage, Rating
24 VAC (RMS)⁄60 VDC
Measurement Range
-29 to149 °C for pipe diameters from 6.4 to 34.9 mm
Models 80pk-1 And 80pj-1 (j-type) Bead Probes. Measurement Range
–40°F to 50°F
Model 80pk-22 Immersion Probe. Measurement Range
–40°F to 1994°F
Model 80pk-3a Surface Probe. Measurement Range
32°F to 500°F
Model 80pk-24 Air Probe. Measurement Range
–40°F to 1500°F
Model 80bk Integrated Dmm Temperature Probe. Measurement Range
–40°F to 500°F
Model 80pk-27 Industrial Surface Probe. Measurement Range
–196°F to 1112°F
Model 80pk-8 Pipe Clamp. Measurement Range
–20°F to 300°F
Equipment Type
Pipe Clamp Temperature Probe
Measures
Temperature
For Use With
Durable Ribbon Sensor, Pipe Surfaces
Lead Free Status / RoHS Status
na
Lead Free Status / RoHS Status
na
®
80PK-8
Type K Thermocouple Pipe Clamp
Temperature Probe
Instruction Sheet
WARNING
To avoid electrical shock, do not use this
probe when voltages exceeding 24V ac or 60V
dc are present. The probe is electrically
connected to the output terminals.
ATTENTION
Pour éviter les risques d’ électrocution, ne pas
utiliser cette sonde quand la tension est
sepérieure à 24 V efficaces en courant
alternatif ou 60 V en courant continu. La pointe
de la sonde est reliée électriquement aux
bornes de sorties.
WARNUNG
Um Elektroschock zu vermeiden, dürfen Sie
den Meßstift nicht in der Anwesenheit von
Spannungen über 24V Wechselstrom oder 60V
Gleichstrom verwenden. Die Spitze des
Meßstifts ist mit den Ausgangsanschlüssen
elektrisch verbunden.
ADVERTENCIA
Para evitar descargas eléctricas, no use esta
sonda cuando se apliquen voltajes que
sobrepasen 24 V ca rcm o 60 V cc. La punta de
la sonda está conectada eléctricamente a las
terminales de salida.
Introduction
The 80PK-8 Type K Thermocouple Pipe Clamp
Temperature Probe (Figure 1) is designed to
reliably measure the temperature of pipes 1/4-inch
(6 mm) to 1 3/8-inch (35 mm) in diameter, at
temperatures between -29 C and 149 C (-20 F
and 300 F), while retaining a long service life.
February 1995 Rev. 1, 4/02
©
1995-2002 Fluke Corporation. All Rights Reserved. Printed in U.S.A.
All product names are trademarks of their respective companies

80PK-8 Summary of contents

Page 1

... V ca rcm cc. La punta de la sonda está conectada eléctricamente a las terminales de salida. Introduction The 80PK-8 Type K Thermocouple Pipe Clamp Temperature Probe (Figure 1) is designed to reliably measure the temperature of pipes 1/4-inch (6 mm 3/8-inch (35 mm) in diameter, at temperatures between -29 C and 149 C (-20 F and 300 F), while retaining a long service life ...

Page 2

... V Notch Figure 1. 80PK-8 Pipe Clamp Temperature Probe Specifications Measurement Range: 300 F) Accuracy: ±1.9 C (±.3.4 F). Surface measurement accuracy of ±1.9 C includes ±1.1 C variation in alloy plus probe contact and shunting uncertainty of ±0.8 C. Repeatability: Output ( 1.00 mV (Reference junction Seebeck Coefficient : @ 40. Measurement Time seconds (5 time constants = 1 complete step change, i ...

Page 3

... If the instrument does not readout properly, refer to “Troubleshooting”. Measurment Considerations The 80PK-8 is designed to be compatible with any temperature-measuring instrument that accepts miniature Type K thermocouples with cold reference junction compensation. The accuracy of the temperature-measurement is a factor of the combined accuracy of the temperature measuring instrument and the 80PK-8 ...

Page 4

... Check for an open circuit indicator on the measuring instrument. Some thermometers have a built in circuit to indicate if the connected probe is open. (All Fluke instruments have this feature.) Refer to instruments owners manual to see if this feature is available. If you suspect a broken connection, use an ohmmeter to check probe continuity from pin to pin ...