KC27 GOSSEN METRAWATT, KC27 Datasheet - Page 5

no-image

KC27

Manufacturer Part Number
KC27
Description
PROBES, KELVIN
Manufacturer
GOSSEN METRAWATT
Series
METRAHITr
Datasheet

Specifications of KC27

Svhc
No SVHC (18-Jun-2010)
Lead Free Status / RoHS Status
na
Fuses
Fuse links for all mΩ
measuring ranges
Acoustic Signal
Electrical Safety
Safety class
Overvoltage
category
Operating voltage
Fouling factor
Test voltage
Electromagnetic Compatibility (EMC)
Interference emission EN 61326: 2002 class B
Interference immunity EN 61326: 2002
Data Interface
With BD232 interface adapter as accessory:
Data transmission
Type
Bidirectional baud rate (read and write)
Ambient Conditions
Accuracy range
Operating temp.
Storage temperature −25 °C ... +70 °C (without batteries)
Relative humidity
Elevation
Deployment
Mechanical Design
Protection
Dimensions
Weight
GOSSEN METRAWATT GMBH
FF (UR) 1.6 A/1000 V AC/DC,
6.3 mm x 32 mm, 10 kA switching
capacity at 1000 V AC /DC and ohmic
load
For display > 610 V in 600 V range
(intermittent tone, 250 ms on/off)
II per IEC/EN 61010-1:2001
/VDE 0411-1:2002
II
600 V
2
3.5 kV~ per IEC/EN 61010-1:2001/
VDE 0411-1:2002
EN 61000-4-2: 1995/A1: 1998
Feature A:
EN 61000-4-3: 1995/A1: 1998
Feature B:
Optical via infrared light through the
housing
RS 232 C, serial, per DIN 19241
SI232-II: all baud rates
BD232: 9600 baud
−10 °C ... +50 °C
40% ... 60%,
no condensation allowed
to 2000 m
Indoors only, except within specified
ambient conditions
Housing: IP 54,
Connector jacks: IP 20
84 mm x 195 mm x 35 mm
Approx. 420 gr. with batteries
(without GH18 protective rubber cover)
Milliohm Resistance Meter and Digital Multimeter,
0 °C ... +40 °C
8 kV atmospheric discharge
4 kV contact discharge
3 V/m
Insulation Tester and Data Logger
Standard Equipment
METRA HIT 27M including
1 GH18 protective rubber cover with carrying strap
3 size AA alkaline manganese batteries
1 KS17S measurement cable set
1 operating instructions
1 abbreviated operating instructions
1 DKD calibration certificate
METRA HIT 27I including
1 GH18 protective rubber cover with carrying strap
3 size AA rechargeable NiMH batteries
1 NA5/600 charging unit
1 KS17S measurement cable set
1 set of Kelvin clips KC4 (1 set = 2 each)
1 abbreviated operating instructions
1 operating instructions
1 DKD calibration certificate
METRA HIT 27AS (avionics set) including
1 METRA HIT 27I
1 GH18 protective rubber cover with carrying strap
3 size AA rechargeable NiMH batteries
1 NA5/600 charging unit
1 KS17S measurement cable set
1 set of Kelvin clips KC4 (1 set = 2 each)
1 set of Kelvin probes KC27 (1 set = 2 each)
1 HC30 hard case
1 abbreviated operating instructions
1 operating instructions
1 BD pack 1 including adapter, cable and METRAwin
1 software METRAwin90 on floppy disk
1 DKD calibration certificate
Accessories
NA5/6000 Charging Unit
Output:
5 V–, max. 600 mA linear controlled, with
low residual ripple and coupling capacity for
mains input; highly insulated (CAT III/600 V)
Input:
230 V~ ±10%; 50/60 Hz
For all countries with a mains voltage
between 220 and 240 V which, how-
ever, have different mains plugs, we rec-
ommend our charging unit in combina-
tion with a commercially available mains
adapter for Euro plugs.
METRAHit
METRA HIT 27M and 27I
software on floppy disk
10/
5