U8202A AGILENT TECHNOLOGIES, U8202A Datasheet - Page 7

Test Leads U3400 U3606A U3401A U3402A

U8202A

Manufacturer Part Number
U8202A
Description
Test Leads U3400 U3606A U3401A U3402A
Manufacturer
AGILENT TECHNOLOGIES
Datasheet

Specifications of U8202A

Voltage Rating
1000V
Kit Contents
Alligator Clips, Mini Grabber Sand Fine-tips Test Probes And Test Probe Leads
For Use With
U3400 Series, U3401A, U3402A, And U3606A
Lead Free Status / RoHS Status
na
Digital multimeter specifications (continued)
Measurement speeds (typical)
1. Reading rate of the A/D converter.
2. Number of measurements per second that can be read through USB using SCPI “READ?” command.
3. Number of measurements per second that can be read through GPIB using SCPI “READ?” command.
Supplementary specifications
Function
DC voltage
(10 V)
DC current
(1 A)
AC voltage
(10 V at 1 kHz)
AC current
(1 A at 1 kHz)
Resistance
(100 kΩ)
Capacitance
(10 µF)
Frequency
(voltage path at
10 V, 1 kHz)
Frequency
(current path at
10 V, 1 kHz)
DC voltage
Measurement method
Maximum input voltage
Input impedance
Input protection
Response time
DC current
Measurement method
Maximum input current
Burden voltage and shunt resistance
Input protection
Response time
Rate
Slow (5½ digits)
Fast (4½ digits)
Slow (5½ digits)
Fast (4½ digits)
Slow (5½ digits)
Fast (4½ digits)
Slow (5½ digits)
Fast (4½ digits)
Slow (5½ digits)
Fast (4½ digits)
Slow/Fast (3½ digits)
Slow (5½ digits)
Fast (4½ digits)
Slow (5½ digit)
Fast (4½ digit)
Sigma Delta A-to-D converter
1000 V
10 MΩ ± 2% range (typical) in parallel with capacitance < 120 pF
1000 V
Approximately 0.15 s when the displayed reading reaches 99.9% DC value of the tested
input signal at the same range
Sigma Delta A-to-D converter
10 mA to 3.0 A DC
• < 0.2 V, 10 Ω for 10 mA range
• < 0.2 V, 1 Ω for 100 mA range
• < 0.3 V, 0.1 Ω for 1 A range
• < 0.7 V, 0.01 Ω for 3 A range
Protected with 3.15 A/500 V, FF fuse
Approximately 0.15 s when the displayed reading reaches 99.9% DC value of the tested
input signal at the same range
dc
rms
Reading speed
(readings/second)
15
70
15
70
15
70
15
70
15
70
5
9
9
9
9
on all ranges
on all ranges, < 0.3 A short circuit
7
1
Reading speed
over USB
(readings/second)
17
31
17
37
17
31
16
37
17
27
4.4
2.7
2.7
2.7
2.7
2
Reading speed
over GPIB
(readings/second)
17
32
17
36
17
32
17
37
17
31
4.6
2.7
2.7
2.7
2.7
3