U2001H AGILENT TECHNOLOGIES, U2001H Datasheet - Page 12

USB POWER SENSOR, 6GHZ, -50DBM To +30DBM

U2001H

Manufacturer Part Number
U2001H
Description
USB POWER SENSOR, 6GHZ, -50DBM To +30DBM
Manufacturer
AGILENT TECHNOLOGIES
Datasheet

Specifications of U2001H

Frequency Measuring Range
10MHz To 6GHz
Amplitude Accuracy
± 4%
Standing Wave Ratio
1.15
Connector Type
N
External Height
36mm
External Width
46mm
External Depth
207mm
Weight
0.324kg
Lead Free Status / RoHS Status
na
Specifications (continued)
Zero and measurement noise
Average only Mode
Normal Mode
1. Condition: (i) 0 to 55 °C and (ii) 95% relative humidity at 40 °C non-condensing.
2. Within one hour after zero set, at a constant temperature, after a 24-hour warm-up of the power sensor.
3. The number of averages at 1 for Normal speed, gate length of 2.27 ms, measured over one minute interval and two standard deviations.
4. The Noise Per Sample specification is only applicable for gated power working range stated in the “Normal Mode Key Specifications and
Power range
U2000/1/2/4A
–60 to –35 dBm
–38 to –15 dBm
–20 to –9 dBm
–11 to –5 dBm
–7 to 15 dBm
10 to 20 dBm
U2000/1/2H sensors
–50 to –25 dBm
–28 to –5 dBm
–10 to 1 dBm
–1 to 5 dBm
3 to 25 dBm
20 to 30 dBm
U2000/1B sensors
–30 to –5 dBm
–8 to 15 dBm
10 to 21 dBm
19 to 25 dBm
23 to 44 dBm
Range
U2000/1/2A
–38 to –15 dBm
–20 to –9 dBm
–11 to –5 dBm
–7 to 15 dBm
10 to 20 dBm
U2000/1/2H
–28 to –5 dBm
–10 to 1 dBm
–1 to 5 dBm
3 to 25 dBm
20 to 30 dBm
U2000/1B
–8 to 15 dBm
10 to 21 dBm
19 to 25 dBm
23 to 44 dBm
Characteristics” table.
1
1
Zero set (internal)
± 1.5 nW (±2.8 nW for U2004A)
± 2 nW (±3 nW for U2004A)
± 12 nW
± 2 µW
± 4 µW
± 6 µW
± 15 nW
± 20 nW
± 120 nW
± 20 µW
± 40 µW
± 60 µW
± 1.8 µW
± 2 µW
± 12 µW
± 2 mW
± 4 mW
Zero set
(internal)
47 nW
530 nW
30 µW
32 µW
270 µW
730 nW
5.3 µW
330 µW
440 µW
3.9 µW
47 nW
530 nW
30 µW
32 µW
Zero set
(external)
43 nW
480 nW
27 µW
30 µW
200 µW
500 nW
4.8 µW
270 µW
300 µW
2.8 mW
43 µW
480 µW
27 mW
34 mW
Zero set (external)
± 600 pW
± 1.5 nW
± 10 nW
± 500 nW
± 1 µW
± 5 µW
± 8 nW
± 20 nW
± 100 nW
± 20 µW
± 30 µW
± 60 µW
± 800 nW
± 2 µW
± 10 µW
± 1 mW
± 2 mW
12
Zero drift
25 nW
230 nW
19 µW
24 µW
110 µW
300 nW
3 µW
190 µW
300 µW
1.1 mW
25 µW
230 µW
19 mW
24 mW
2
Zero drift
200 pW
400 pW
1.5 nW
50 nW
500 nW
2 µW
2 nW
4 nW
15 nW
500 nW
5 µW
20 µW
200 nW
400 nW
1.5 µW
50 nW
500 µW
Measurement
noise
28 nW
300 nW
20 µW
21 µW
180 µW
310 nW
5 µW
230 µW
260 µW
2.8 mW
28 µW
300 µW
20 mW
21 mW
3
2
Measurement noise
1 nW
1.5 nW
15 nW
650 nW
1 µW
10 µW
10 nW
15 nW
150 nW
6.5 µW
10 µW
100 µW
1 µW
1.5 µW
15 µW
650 µW
1 mW
Noise per
sample
90 nW
1 µW
55 µW
85 µW
550 µW
900 nW
10 µW
550 µW
850 µW
5.5 mW
90 µW
1 mW
55 mW
85 mW
4
3