Page 1
... Offering the same functionality as the U1252A/ U1252B, the U1253A/U1253B is the world’s fi rst OLED handheld DMM. You won’t have to squint to be sure you’re reading it right: On the the bench, you’ ...
Page 2
... Broad range of measurement functions, including temperature and capacitance Built-in battery charging for optimum capacity [1] U1253A/U1253B only [2] U1252A/U1252B/U1253A/U1253B only 50,000 count dual display with true- RMS reading capability Figure 2: U1253B front view 2 20 MHz frequency 2 counter Data logging to internal or external memory ...
Page 3
... Input impedance: Refer to Table A on page 9. [2] The accuracy could be 0.05 for U1251A and 0.05 for U1252A/3A. Always use the NULL function to zero out the thermal effect before measuring the signal. [3] The accuracy of 500 W and specifi ed after NULL function, which is used to subtract the test lead resistance and the thermal effect. ...
Page 4
... V 500.00 V 1000.0 V U1251A/U1251B AC CURRENT SPECIFICATIONS FUNCTION RANGE [4] 500.00 µA 5000.0 µA 50.000 mA [2] AC CURRENT 440.00 mA 5.0000 A 10.000 A [5] U1252A/U1253A AC VOLTAGE SPECIFICATIONS FUNCTION RANGE 50.000 mV 500.00 mV 1000.0 mV TRUE RMS AC 5.0000 V VOLTAGE [1] [2] [9] 50.000 V 500.00 V 1000.0 V U1252B/U1253B AC VOLTAGE SPECIFICATIONS FUNCTION RANGE 50.000 mV 500 ...
Page 5
... AC mV/V and AC µA/mA/A specifi cations are true RMS AC coupled, valid from 5% to 100% of range. The crest factor may full scale half- scale, except for 1000 mV and 1000 V ranges, where the crest factor is 1.5 at full scale and 3 at half scale. [3] For U1251A / U1252A / U1253A, the input signal is lower than 20,000,000 V-Hz (the product of voltage and frequency). [4] Input current >35 µArms. ...
Page 6
... U1252A/U1252B/U1253A/U1253B AC+DC CURRENT SPECIFICATIONS FUNCTION RANGE [6] 500.00 µA 5000.0 µA 50.000 mA TRUE RMS AC CURRENT [10] 440.00 mA 5.0000 A 10.000 A [7] U1251A/U1251B/U1252A/U1252B TEMPERATURE SPECIFICATIONS THERMOCOUPLE RANGE TYPE ºC K –200 ~ 1372 ºC J –210 ~ 1200 ºC (for U1252A/U1252B) U1253A/U1253B TEMPERATURE SPECIFICATIONS THERMOCOUPLE RANGE TYPE ºC –200 ~ –40 °C K – ...
Page 7
... The additional error to be added as frequency >20 kHz and signal input <10% of range: 3 counts of LSD per kHz. [4] The additional error to be added as frequency >15 kHz and signal input <10% of range: 3 counts of LSD per kHz. [5] For U1252A / U1253A, the input signal is lower than 20,000,000 V-Hz (the product of voltage and frequency). [6] Input current >35 µArms. ...
Page 8
... DUTY CYCLE AND PULSE WIDTH FUNCTION DUTY CYCLE PULSE WIDTH [1] The positive or negative pulse width must be greater than 10 µs, and the duty cycle range should be considered. The pulse width range is determined by the frequency of the signal. U1252A/U1252B/U1253A/U1253B FREQUENCY COUNTER SPECIFICATIONS DIVISION RANGE 1 99.999 Hz (secondary display “ ...
Page 9
... (nominal (nominal), with 1 parallel at dual display > > 1.1 MW (nominal) in 1.1 MW (nominal) in parallel with < 100pF parallel with < 100pF N/A > 1.1 MW (nominal) in parallel with < 100 pF 9 [5] U1252A U1253A 100 100 200 1000 DEFAULT REFERENCE 1-9999 U1251B U1252B/U1253B (nominal (nominal), with parallel at with ...
Page 10
... Group 1, Class A COMMON MODE REJECTION RATIO (CMRR) U1251A/U1251B/U1252A/U1252B: > DC, 50/60 Hz ± 0. unbalanced) U1253A/U1253B: >100 dB at DC, 50/60 Hz ± 0. unbalanced) NORMAL MODE REJECTION RATIO (NMRR) U1251A/U1251B/U1252A/U1252B: > DC, 50/60 Hz ± 0.1% U1253A/U1253B: > DC, 50/60 Hz ± 0.1% TEMPERATURE COEFFICIENT o o 0.15 * (specifi ed accuracy)/ C (from 20 ...
Page 11
... Test leads: CAT III 1000 • Alligator clips: CAT III 1000 • Fine-tip test probes: CAT II 300 • SMT grabbers: CAT II 300 • Mini grabber: CAT II 300 U1252B U1253A U1253B APPLICABLE MODELS U1251A U1252A U1253A • • • • – • • – ...
Page 12
U1163A SMT grabbers • One pair of SMT grabbers (red and black). Recommended for use with Agilent standard test leads. • Rated CAT II 300 MEASURING ACCESSORIES (NON-TEMPERATURE) U1168A Standard test lead kit Includes two test leads ...
Page 13
MEASURING ACCESSORIES (TEMPERATURE) U1186A K-type thermocouple and adapter • T/C adapter J/K-type • T/C bead J-type: –20 °C to 200 °C CABLE U1173A IR-to-USB cable • For remote control and data logging to PC • Max. baud rate: 19,200 bits ...
Page 14
... Spain 34 (91) 631 3300 Sweden 0200- Switzerland 0800 United Kingdom 44 (0) 118 9276201 Other European Countries: www.agilent.com/find/contactus Revised: October 1, 2009 Product specifications and descriptions in this document subject to change without notice. © Agilent Technologies, Inc. 2010 Printed in USA, May 19, 2010 5989-5509EN ...