Specifications of P01196787

Dmm Type
Handheld
Voltage Measuring Range Dc
50mV-1000V
Voltage Measuring Range Ac
50mV-1000V
Current Range Dc
500 µA - 10A
Current Range Ac
500 µA - 10A
Resistance Measuring Range
500
Frequency Measuring Range
99.999Hz To 999.99kHz
Rohs Compliant
NA
Lead Free Status / RoHS Status
na
n
n
n
• Crest factor ≤ 3
n
n
n
50,000-count TRMS AC/DC multimeter
(1) Protection 1,000 Vrms for Ip-p < 0.3 A
(8) An additional 0.5% error should be applied to
the accuracy of the 10 A calibre when measuring
currents between 10 A and 20 A for a maximum
of 30 seconds. After a current measurement >
10 A, and before carrying out any low-current
measurements, leave the multimeter to cool for
twice the time that the current was applied.
(10) Minimum intensity measured Imin >35 µArms
500 nS
500 µA
50 MΩ
1,000 nF
1,000 µF
500 Ω
Crest value (capture of changes)
DC Current
AC Current (root mean square value: 5 % to 100 % of the range)
Resistance
Diode test
Capacitance
5 000 µA
100 mF
5 kΩ
Diode
Range
100 nF
100 µF
500 mA
10 mF
500 kΩ
Range
10 A
Range
Range
10 nF
10 µF
50 mA
50 kΩ
5 000 µA
5 MΩ
500 mA
Range
500 µA
50 mA
10 A
5 A
5 A
Unique event > 1 ms
Repetitive > 250 µs
(23)
(8)
(23)
(1)
(24)
(10)
(25)
(8)
Signal width
Resolution
0.001 mF
Resolution
0.001 nF
0.001 µF
0.01 mF
Resolution
Resolution
0.0001 MΩ
0.01 nF
0.01 µF
0.0001 kΩ
0.001 MΩ
0.1 nF
0.1 µF
0.001 mA
0.001 kΩ
(1)
0.0001 A
0.1 mV
0.01 mA
0.01 kΩ
0.01 µA
0.001 A
0.01 nS
0.01 Ω
0.1 µA
Resolution
(1)
/ Continuity test
0.001 mA
0.0001 A
0.01 mA
0.01 µA
0.001 A
0.1 µA
Accuracy
3 % + 10
1 % + 8
1 % + 5
2 % + 40
2 % + 40
0.08 % + 10
1.5 % + 50
1.5 % + 40
1.5 % + 40
1.5 % + 40
0.05 % + 5
0.08 % + 5
30~45 Hz
Accuracy
Accuracy
0.2 % + 5
1 % + 10
1 % + 5
0.1 % + 5
0.1 % + 5
0.2 % + 5
0.2 % + 5
0.2 % + 10
Accuracy
0.2 % + 5
Accuracy for mV / Voltage/DC Current
(22)
(22)
(11 For currents lower than 3 Arms, add 50 counts: 2.2 %+ 50
(12) For currents lower than 3 Arms, add 45 counts: 2.2 % + 45
(21) Always use the “Relative” function to offset the thermal effect with the
measurement leads open before measuring the signal. If the “Relative” function is not
used, you must add 20 counts to the accuracy. The thermal effect could occur in the
following cases:
- Overheating due to manipulation error by a applying a high voltage of 50 V to
- Overheating after battery recharging.
- Overheating after measuring a current higher than 500 mA: it is recommended to
1,000 V for resistance measurement, diode test and mV functions.
leave the multimeter to cool for twice the time that the current is applied.
2 % + 1,000 for all the ranges
2 % + 400 for all the ranges
Measurement current
Full-scale measurement rate
Measurement current
0.8 % + 20
0.8 % + 20
0.8 % + 20
0.8 % + 20
0.8 % + 20
0.8 % + 20
(21)
(21)
(21)
(21)
45~2 kHz
Accuracy
Approx. 1.0 mA
0.38 mA
0.01 times/sec.
1,0 mA
345 nA
200 nA
200 nA
0.1 times/sec.
3.8 µA
38 µA
4 times/sec.
1 time/sec.
Load voltage / Shunt
< 3 A / 5 kHz
0.06 V (100 Ω)
0.2 V (0.01 Ω)
0.4 V (0.01 Ω)
0.6 V (100 Ω)
2 k~20 kHz
0.09 V (1 Ω)
3 % + 80
3 % + 60
3 % + 60
3 % + 60
3 % + 60
0,9 V (1 Ω)
Protection against overloads
Open circuit voltage
11 A HRC fuse, 10 x 38 mm
1,000 Vrms
Protection against
< + 4.8 V
440 mA HRC fuse
1,000 V / 30 kA
1,000 V / 30 kA
For capacitors with low values, use the “Relative” mode to offset the
residual capacitance.
The maximum display for each range is 11,000 counts.
10 x 38 mm
overloads
dc
(1)
Protection against overloads
440 mA HRC fuse
1,000 V / 30 kA
1,000 V / 30 kA
11 A HRC fuse
10 x 38 mm
10 x 38 mm
(22) For currents < 3 Arms, add un uncertainty of 40
counts (2 % + 40)
(23) The accuracy of 500 Ω and 5 kΩ is specified after
application of the relative function, which is used to
subtract the resistance of the measurement leads and
the thermal effect.
(24) For the 50 MΩ range, the relative humidity is
specified as less than 60 %.
(25) The accuracy is specified for < 50 nS, after
application of the relative function when the
measurement leads are open.
General remarks:
Maximum open-circuit voltage: < + 4.8 V
Instantaneous continuity: the built-in buzzer
sounds when the resistance is less than 10.0 Ω.
The built-in buzzer is triggered when the reading
is below 50 mV approx.