34405A AGILENT TECHNOLOGIES, 34405A Datasheet - Page 4

MULTIMETER, DIGITAL, BENCH, 5-1/2 DIGIT

34405A

Manufacturer Part Number
34405A
Description
MULTIMETER, DIGITAL, BENCH, 5-1/2 DIGIT
Manufacturer
AGILENT TECHNOLOGIES
Datasheet

Specifications of 34405A

Dmm Type
Bench
Voltage Measuring Range Dc
100mV To 1000V
Voltage Measuring Range Ac
100mV To 750V
Current Measuring Range Dc
10mA To 10A
Current Measuring Range Ac
10mA To 10A
Lead Free Status / RoHS Status
na
AC CHARACTERISTICS
TEMPERATURE and CAPACITANCE CHARACTERISTICS
[1] Specifi cations are for 30 minutes warm-up, 5 1/2 digit resolution and calibration temperature 18 °C - 28 °C.
[2] 20% over range on all ranges except 1000 Vdc.
[3] Specifi cations are 2-wire ohms using Math Null. If without Math Null, add 0.2 Ω additional error.
[4] Specifi cations are for the voltage measured at the input terminals only.
[5] 20% over range on all range except 750 Vac
[6] Specifi cations are for sinewave inputs >5% of range. Maximum crest factor : 3 at full scale.
[7] Additional error to be added as frequency >30 kHz and signal input <10% of range. 30 kHz ~ 100 kHz: 0.003% of full scale
[8] For 12 A terminal, 10 A dc or ac rms continuous, >10 A dc or ac rms for 30 seconds ON and 30 seconds OFF.
[9] For 1 A and 10 A ranges, the frequency is verifi ed for less than 5 kHz.
[10] Specifi cations are for half-hour warm-up, using 0.1 second aperture. The frequency can be measured up 1 MHz as 0.5 V
[11] For 20 Hz ~ 10 kHz, the sensitivity is AC input current from 10% to 120% of range except where noted.
[12] For 100 kHz ~ 300 kHz, the sensitivity will be 12% ~ 120% of range except 750 V range.
[13] Input Impedence is in paralleled with capacitance <120 pF.
[14] For input <200 V rms
[15] For input <300 V rms
[16] For frequency, use AC Accuracy ± (% of reading + 3 counts)
[17] Minimum measured frequency is 1 Hz
FUNCTION
FREQUENCY
FUNCTION
TEMPERATURE
CAPACITANCE
per kHz.
signal to 100 mV/1 V ranges.
[10],[16]
RANGE
100.000 mV to 750.00 V
10.0000 mA to 10.0000 A < 2 Hz
RANGE
-80 °C - 150 °C
-110.0 °F - 300.0 °F
1.000 nF
10.00 nF
100.0 nF
1.000 μF - 100.0 μF
1000 μF
10 000 μF
[5]
[1]
TEST CURRENT, etc.
5 kΩ thermistor probe
5 kΩ thermistor probe
0.75 μA
0.75 μA
8.3 μA
83 μA
0.83 mA
0.83 mA
FREQUENCY
< 2 Hz
< 20 Hz
20 Hz - 100 kHz
100 kHz ~ 300 kHz
< 20 Hz
20 Hz ~ 10 kHz
[17]
[17]
4
[11]
[11]
[12]
1 Year
23 °C ± 5 °C
Probe accuracy + 0.2 °C
Probe accuracy + 0.4 °F
2.0+0.8
1.0+0.5
1.0+0.5
1.0+0.5
1.0+0.5
2.0+0.5
[1]
ACCURACY ± (% of reading + % of range)
ACCURACY ± (% of reading + % of range)
1 Year
23 °C ± 5 °C
0.18+0.003
0.04+0.003
0.02+0.003
0.02+0.003
0.18+0.003
0.04+0.003
0.02+0.003
Temperature Ceffi cient
0 °C - 18 °C
28 °C - 55 °C
0.005
0.005
0.005
0.005
0.005
0.005
0.005
Temperature Ceffi cient
0 °C - 18 °C
28 °C - 55 °C
0.002 °C
0.0036 °F
0.02+0.001
0.02+0.001
0.02+0.001
0.02+0.001
0.02+0.001
0.02+0.001