5360 B&K Precision, 5360 Datasheet

DMM TRU RMS 50K COUNT

5360

Manufacturer Part Number
5360
Description
DMM TRU RMS 50K COUNT
Manufacturer
B&K Precision
Type
Digital (DMM)r
Datasheets

Specifications of 5360

Includes
Battery, Test Leads
Style
Handheld
Display Digits
5
Display Type
LCD, Bar Graph
Display Count
50000
Function
Voltage, Current, Resistance, Capacitance, Frequency
Functions, Extra
Continuity, Diode Test
Features
Auto Off, RS-233 Port
Ranging
Auto/Manual
Response
True RMS
Accuracy
+/- 0.1 %
Lead Free Status / Rohs Status
Lead free / RoHS Compliant
Other names
BK-5360
BK5360
AC VOLTAGE
(Auto/Manual Ranging, True RMS)
Modes: AC or AC+DC, selectable
Accuracy at greater than 10% of range, 50,000 counts at Full Scale:
Ranges
500 mV
5V
50V
500V
750V
50 kHz to 100 kHz (5V, 50 V, 500 V ranges): ±6% FS typical
Crest Factor: 6
Accuracy at greater than 10% of range, 5,000 counts at Full Scale:
Ranges
500 mV
5V
50V
500V
750V
Crest Factor: 3
28
Specifications
The 5300 Series handheld DMM is the most advanced digital mul-
timeter available in this price range. These meters incorporate a
special 50,000 count integrated circuit for high-accuracy (0.025%)
basic accuracy, high-reliability design. Measure AC Volts (True RMS
AC, AC + DC), Resistance, Capacitance and Frequency, in addition
to dB, Duty Cycle, Pulse Width, Power Line Disturbances and
Temperature. These meters also feature fuse and battery access
without breaking a calibration seal, great for ISO facilities. All mod-
els offer Peak Hold, Hold, Relative Mode and Backlight.
Precision 50,000 count instruments
0.025% Basic DCV accuracy (Models 5380 and 5390)
True RMS reading
Large LCD display provides easy readability, all test
parameters in clear view.
Best value of any DMM
CE safety approved
0.3%R+30D
0.3%R+30D
0.3%R+30D
0.3%R+30D
0.3%R+30D
DC to
1kHz
DC to 1kHz
1%R+3D
1%R+3D
1%R+3D
1%R+3D
1%R+3D
5360
5390
1%R+30D
1%R+30D
1%R+3D
1%R+3D
4%R+3D
1%R+3D
1%R+3D
1%R+3D
1 kHz to
1 kHz to
5 kHz
4 kHz
5380
2%R+30D
2%R+3D
2%R+3D
5 kHz to
10 kHz
5390
Not specified
5%R+30D
2%R+30D
2%R+30D
2%R+30D
4 kHz to
10 kHz
Basic Functions
Frequency
Temperature Probe
Duty Cycle
Event Counter
Backlight
Basic Accuracy
Voltage (Best Resolution)
Current (Best Resolution)
5370
Features
2%R+30D
2%R+30D
2%R+30D
10 kHz to
Not specified
Not specified
30 kHz
2%R+3D
2%R+3D
2%R+3D
10kHz to
20 kHz
5380
Not specified
3%R+30D10 mV
3%R+30D1 mV
3%R+30D
30 kHz to
models
5390
50 kHz
model
20 kHz to
3%R+3D
3%R+3D
3%R+3D
30 kHz
0.025%
750VAC/1000VDC
10AAC/DC
5370
5380
(10µV)
(10nA)
10 M /100 pF
10 M /100 pF
Resolution
Resolution
100 µV
100 mV
100 mV
100µV
10m
10µV
1m V
1 V
5370
0.05%
10 M /100 pF
11 M /100 pF
10 M /100 pF
10 M /100 pF
10 M /100pF
750VAC/1000VDC
Impedance
models
5360
Input
10AAC/DC
10 M /100 pF
11 M /100 pF
10 M /100pF
5360
(100µV)
Impedance
0.1%
(1µA)
or 1 G
Input

5360 Summary of contents

Page 1

... Not specified 5 kHz to 10kHz to 10 kHz 20 kHz Not specified 2%R+30D 2%R+3D 2%R+3D 2%R+3D 2%R+3D 2%R+3D Not specified 5370 5360 models 5380 5370 5360 0.025% 0.05% 0.1% 750VAC/1000VDC 750VAC/1000VDC (10µV) (100µV) 10AAC/DC 10AAC/DC (10nA) (1µA) Input Resolution Impedance 10µ /100 100 µV ...

Page 2

... MEASUREMENT - 5390 and 5370 - Measures value relative to a reference from 1 to 9999 , 5380 at 600 PEAK HOLD - Selection: Peak + or Peak- (surge or sag) Acq. Time: <1rms DUTY CYCLE: 5390, 5370, 5360-Modes: Selectable %+ ,or %- Min EVENT COUNTER (Model 5390) - Range 99999 (Totalize) MIN PULSE DURATION: 2 µs PULSE WIDTH (Model 5390) Modes: Selectable +, - Min. Pulse Width: 20µ ...