FLUKE-113 Fluke Electronics, FLUKE-113 Datasheet - Page 2

no-image

FLUKE-113

Manufacturer Part Number
FLUKE-113
Description
MULTIMETER UTILITY TRU-RMS 600V
Manufacturer
Fluke Electronics
Series
11xr
Type
Digital (DMM)r
Datasheets

Specifications of FLUKE-113

Includes
Battery, Test Leads
Style
Handheld
Display Digits
3.75
Display Type
LCD, Bar Graph
Display Count
6000
Function
Voltage, Resistance, Capacitance
Functions, Extra
Continuity, Diode Test
Features
Backlight, Hold, Min/Max, Sleep
Ranging
Auto/Manual
Response
True RMS
Lead Free Status / RoHS Status
Lead free / RoHS compliant by exemption
Other names
3078875
Accuracy is specified for 1 year after calibration, at operating temperatures of 18 °C to 28 °C,
with relative humidity at 0 % to 95 %. Accuracy specifications are given as follows:
Note:
Specifications are subject to change without notice.
± ( [ % of reading ] + [ number of least significant digits ] )
Notes:
1
2
3
2 Fluke Corporation
Function
Function
All
specified, it is possible and normal for this and other true-rms meters to display non-zero readings when the test
leads are disconnected from a circuit or are shorted together.
Crest factor of ≤ 3 at 4000 counts, decreasing linearly to 1.5 at full scale.
After measuring voltage, a wait time of 1 minute is required to maintain accuracy of ohms, capacitance, diode test,
and continuity.
Maximum voltage between any
terminal and earth ground
Display
Operating temperature
Storage temperature
Temperature coefficient
Operating altitude
Storage altitude
Relative humidity
Battery type
Battery life
Shock
Vibration
Size (HxWxL)
Weight
Safety
EMI regulations
Function
Chek
3
3
3
3
Chek
Chek voltage ranges are specified from 60 counts to 100 % of range. Because inputs below 60 counts are not
1,2
< 2.7 V dc
< 2.7 V dc
Open Circuit Test Voltage
Range
6.000 V
60.00 V
600.0 V
Range
2.000 V
600.0 Ω
6.000 kΩ
60.00 kΩ
1000 nF
10.00 µF
100.0 µF
9999 µF
Common Mode Rejection Ratio (1 kΩ Unbalanced)
113 Utility Meter
Specifications
Resolution
0.001 V
0.01 V
0.1 V
Resolution
0.001 V
0.1 Ω
0.001 kΩ
0.01 kΩ
1 nF
0.01 µF
0.1 µF
1 µF
Input Impedance (Nominal)
> 60 dB at dc, 50 or 60 Hz
Short Circuit Current
600 V
3 3/4-digits, 6000 counts, updates 4/sec
-10 °C to 50 °C (14 °F to 122 °F)
-40 °C to 60 °C (-22 °F to 140 °F)
0.1 x (specified accuracy)/°C (< 18 °C or > 28 °C)
2,000 m
10,000 m
95 % to 30 °C
75 % to 40 °C
45 % to 50 °C
9 Volt Alkaline, ANSI 1604A / IEC 6F22
Alkaline: 300 hours typical, without backlight
1 Meter drop per IEC 61010-1-2001
Per MIL-PRF-28800 for Class 2 instrument
167.1 mm X 85.1 mm X 46.0 mm (6.58 in X 3.35 in X 1.81 in)
13.0 oz (404 g)
Complies with ANSI/ISA 82.02.01 (61010-1) 2004, CAN/CSA-C22.2 No
61010-1-04, UL 61010-1 (2004) and IEC/EN 61010-1 2nd Edition for
measurement Category III, 600 V, Pollution Degree 2, EMC EN61326-1
Complies with FCC Part 15, Class B
~3 kΩ < 300 pF
< 350 µA
< 1.0 mA
1.9 % + 2
1.9 % + 2
1.9 % + 2
100 µF - 1000 µF: 1.9 % + 2
> 1000 µF: 5 % + 20
DC, 45 to 500 Hz
Beeper on < 20 Ω, off > 250 Ω; detects
opens or shorts of 500 µs or longer
2.0 % + 3
Full Scale Voltage
2.000 V dc
< 0.7 V dc
2.0 % + 3
0.9 % + 2
0.9 % + 1
0.9 % + 1
Accuracy
Accuracy
500 Hz to 1 kHz
4.0 % + 3
Fluke. Keeping your world
Fluke Corporation
PO Box 9090, Everett, WA USA 98206
Fluke Europe B.V.
PO Box 1186, 5602 BD
Eindhoven, The Netherlands
For more information call:
In the U.S.A. (800) 443-5853 or
Fax (425) 446-5116
In Europe/M-East/Africa (31 40) 2 675 200 or
Fax (31 40) 2 675 222
In Canada (800)-36-FLUKE or
Fax (905) 890-6866
From other countries +1 (425) 446-5500 or
Fax +1 (425) 446-5116
Web access: http://www.fluke.com
©2008 Fluke Corporation. Specifications subject
to change without notice. Printed in U.S.A.
3/2008 3292015 D-EN-N Rev A
Ordering
information
Fluke 113 Utility Meter
Included
Fluke 113 Utility Multimeter
TL75 Test Leads
One 9V Battery
Printed Users Manual
up and running.
®