U1571A AGILENT TECHNOLOGIES, U1571A Datasheet - Page 4

NIMH BATTERY, 7.2V, 4.5AH

U1571A

Manufacturer Part Number
U1571A
Description
NIMH BATTERY, 7.2V, 4.5AH
Manufacturer
AGILENT TECHNOLOGIES
Datasheet

Specifications of U1571A

Battery Capacity
4500mAh
Battery Voltage
7.2V
Battery Technology
Nickel Metal Hydride
Battery Terminals
Connector
For Use With
Agilent Technologies U1602A And U1604A Hand-Held Digital Oscilloscopes
Lead Free Status / RoHS Status
na
SCOPE SPECIFICATIONS
SCOPE CHARACTERISTICS
[1] All specifications are warranted. Specifications are valid after a 30-minute warm-up period and within a range of ±10 ºC from
[2] All characteristics are typical performance values and are not warranted. Characteristics are valid after a 30-minute warm-up
[3] Maximum sampling rates are shown here. Sampling rate will vary according to selected time base. For more information, please refer
[4] 200 MS/s sampling rate is only available at 125 ns/div timebase
[5] Number of points displayed will vary according to time base selected. For more information, refer to User’s and Service Guide
Vertical System: Scope Channels
Bandwidth (–3 dB)
DC vertical gain accuracy
Scope Channel Triggering
Trigger sensitivity
Acquisition: Scope Channels
Maximum sample rate
Equivalent sample rate
Vertical resolution
Maximum recording length:
Peak detection
Average
Vertical System: Scope Channels
Analog channels
Bandwidth (–3 dB)
AC coupled
Rise time
Single shot bandwidth
Vertical sensitivity
Maximum input
Offset/Dynamic range
Input impedance
Coupling
firmware calibration temperature.
period and within a range of ±10 ºC from firmware calibration temperature.
to User’s and Service Guide.
[1]
U1602A/U1602B: DC to 20 MHz
U1604A/U1604B: DC to 40 MHz
5 mV/div to 20 mV/div: ± 5% full scale
50 mV/div to 100 V/div: ± 3% full scale
DC to 5 MHz: 0.8 divisions
U1602A/U1602B: 5 MHz to 20 MHz – 1 division
U1604A/U1604B: 5 MHz to 40 MHz – 1 division
100 MSa/s per channel (50 s/div to 250 ns/div)
200 MSa/s single channel and interleaved (125 ns/div)
U1604A/U1604B: 2.5 GSa/s (125 ns/div to 10 ns/div)
8 bits
U1602A/U1604A: Up to 11 100 points, viewable on screen with zoom function
U1602B/U1604B: Up to 125 000 points, viewable on screen with zoom function
5 ns
Selectable in average number of 2, 4, 8, 16, 32, 64, 128, 256
Channel 1 and Channel 2 simultaneous acquisition
U1602A/U1602B: DC to 20 MHz
U1604A/U1604B: DC to 40 MHz
< 10 Hz without probe
< 1 Hz with 10 MW 10:1 probe
U1602A/U1602B: < 17.5 ns
U1604A/U1604B: < 8.8 ns
U1602A/U1602B: 20 MHz
U1604A/U1604B: 40 MHz
5 mV/div to 100 V/div (1:1 scope probe)
50 mV/div to 1 kV/div (10:1 scope probe)
500 mV/div to 10 kV/div (100:1 scope probe)
CAT III 300 Vrms (up to 400 Hz) from terminal to ground
± 5 div
1 MW I I < 20 pF
AC, DC, GND
[2]
4
3
4

Related parts for U1571A