EC400 EXTECH INSTRUMENTS, EC400 Datasheet

METER, CONDUCTIVITY/SALINITY 19.99mS/CM

EC400

Manufacturer Part Number
EC400
Description
METER, CONDUCTIVITY/SALINITY 19.99mS/CM
Manufacturer
EXTECH INSTRUMENTS
Datasheet

Specifications of EC400

Conductivity Range
0mS/cm To 19.99mS/cm
Tds Range Max
999ppm
Salinity Range Max
999ppm
Temperature Measuring Range
0°C To 65°C
Lead Free Status / RoHS Status
na
www.extech.com
ExStik
Ordering Information:
EC400 ............ExStik
EC410 ............ExStik
EC405 ............Spare Conductivity Cell
EX006 ............Weighted base with 5 solution cups
EX007 ............Spare sample solution cups (24pk)
EC-84-P..........84μS Conductivity Standard (2 bottles - 1 pint each)
EC-1413-P......1413μS Conductivity Standard (2 bottles - 1 pint each)
EC-12880-P....12880μS Conductivity Standard (2 bottles - 1 pint each)
CA895 ............Small vinyl carrying case with belt loop
CA903 ............Carrying case for ExStik
EX010 ............Extension Cable, 3ft (1 Meter) with probe weight for ExStik
EX050 ............Extension Cable, 16ft (5 Meters) with probe weight for ExStik
Features:
Autoranging meter offers 3 ranges of measurements from tap water to
wastewater and any aqueous solution requiring a conductivity,
TDS or Salinity measurements
Units of measure include μS/cm, mS/cm, ppm, ppt, mg/L, and g/L
Adjustable Conductivity to TDS ratio factor from 0.4 to 1.0 —
conveniently calculates the TDS value
Large 3-1/2 digit (2000 count) Digital display with
analog bargraph to indicate sample trends
Data Hold, Auto Power Off and low battery indication
Automatic Temperature Compensation of 2% per °C
Simultaneous display of Conductivity, TDS, or Salinity plus Temperature
Internal memory stores up to 25 labeled readings for easy recall
Self calibration of electronics on power up
Easy to replace Conductivity cell module
Waterproof design to withstand wet environment – meets IP57 standard
EC400 includes meter and Conductivity cell*, protective sensor cap,
sample cup with cap, four CR2032 button batteries, and 48"(1.2m)
neckstrap. Order Conductivity standards separately
Optional EC410 kit includes EC400 ExStik® II
Conductivity/TDS/Salinity/Temperature Meter, with sensor, protective
cap, 84μS/cm, 1413μS/cm and 12880μS/cm calibration standards
(30mL bottle each), weighted base, 3 sample plastic cups with caps,
four 3V CR-2032 button batteries, 48" (1.2m) neckstrap,
and carrying case
Conductivity, TDS, Salinity, and Temperature
that measures four parameters
Waterproof meter with one high accuracy multi-range sensor
®
®
®
Conductivity/TDS/Salinity Meter
Conductivity/TDS/Salinity Kit
II Conductivity/TDS/Salinity Meter
kit
Specifications subject to change without notice.
Copyright © 2007-2008 Extech Instruments Corporation. All rights reserved including the right of reproduction in whole or in part in any form.
Specifications
Conductivity
TDS
Salinity
Temperature
Power
Dimensions
Weight
Ranges
0 to 199.9μS/cm
200 to 1999μS/cm
2.00 to 19.99mS/cm
0 to 99.9ppm (mg/L)
100 to 999ppm (mg/L)
1.00 to 9.99ppt (g/L)
0 to 99.9ppm (mg/L)
100 to 999ppm (mg/L)
1.00 to 9.99ppt (g/L)
32° to 149°F (0 to 65°C)
Four CR2032 button batteries
1.4 x 6.8 x 1.6" (36 x 173 x 41mm)
3.8oz (110g)
EC410 Kit
Max. Resolution
0.1μS/cm
1μS/cm
0.01mS/cm
0.1ppm (mg/L)
1ppm (mg/L)
0.01ppt (g/L)
0.1ppm (mg/L)
1ppm (mg/L)
0.01ppt (g/L)
0.1°F/°C
Basic Accuracy
±2% FS
±2% FS
±2% FS
±1.8°F/1°C