GX320 METRIX, GX320 Datasheet - Page 2

no-image

GX320

Manufacturer Part Number
GX320
Description
FUNCTION GENERATOR, DDS 20MHZ
Manufacturer
METRIX
Datasheets

Specifications of GX320

Signal Generator Type
Signal
Bandwidth
20MHz
Modulation Type
AM, FM
Supply Voltage Range
207V To 253V
External Height
116mm
External Width
180mm
External Depth
227mm
Svhc
No SVHC (15
Lead Free Status / RoHS Status
na
Technical specifications
MAN-MACHINE INTERFACE
Display
Controls on front panel
Adjustment of signal parameters
BNC output terminals on front panel
BNC input terminals on front panel
Continuous signal generation
Frequency
Frequency range
Resolution / accuracy
Long-term drift
Temperature coefficient
Amplitude
Voltage levels
Flatness
Vdc offset
Impedance / Protection
Signal waveform
Sine
Triangle (max. frequency 2 MHz)
Square & "LOGIC"
TTL output
Frequency sweep
Modes
"INT" internal sweep
"EXT" external sweep
Modulation (GX320 only)
Internal AM modulation
External AM modulation
Internal FM modulation
External FM modulation
SHIFT K function (GX320 only)
Internal FSK
External FSK
Internal PSK
External PSK
Burst function (GX320 only)
Internal BURST
External BURST
Gate function (GX320 only)
External GATE
Synchro function (GX320 only)
Set-up of several GX320 in cascade
External frequencymeter
Measurement range
Accuracy
Sensitivity
Input impedance
Safety / Max. measurable voltage
General specifications
Configuration memories (GX320)
Communication interface
Mains power supply
Safety / EMC
Mechanical specifications
Warranty / Origin
References for ordering
GX310: 10 MHz function generator
GX310 P: 10 MHz programmable function generator
GX320: 20 MHz function generator
GX320-P: 20 MHz programmable function generator
France
Chauvin Arnoux
190, rue Championnet
75876 PARIS Cedex 18
Tel : +33 1 44 85 44 58
Fax : +33 1 46 27 07 48
info@metrix.fr
www.metrix.fr
GX310 and GX320 laboratory function generators
UNITED KINGDOM
Chauvin Arnoux Ltd
Waldeck House -Waldeck Road
MAIDENHEAD SL6 8BR
Tel: +44 1628 788 888
Fax: +44 1628 628 099
info@chauvin-arnoux.co.uk
www.chauvin-arnoux.com
Phase switching by a TTL signal with a frequency <1 MHz (VCG IN input) – adjustable over ± 180° (resolution 1° )
1 mV to 20.0 Vpp with open circuit in 3 automatic ranges – 3-digit Vpp or Vrms display – Max. resolution 1 mV
<5% pour 1 mHz< F <10 MHz , and ± 0.5 dB typically up to 20 MHz (GX320) (for a level of 0.1 Vpp to 20 Vpp)
Maximum frequency of signals generated 100 kHz – Adjustment of phase shift over ± 180° (resolution 1 ° )
Modulation by a sine signal with a frequency of 1 kHz – Unlimited excursion between "F Start" & "F Stop"
19 direct-access controls (9 backlit and adjustable) – 1 Main Out On/Off key – 1 digital encoder wheel
Validation of the AC component of "Main Out" by a TTL signal with a frequency <2 MHz (GATE IN input)
0.001 Hz to 10.000 MHz (10 ranges)
Switching between "F Start" & "F Stop" by a TTL signal with a max. frequency of 1 MHz (VCG IN input)
Continuous by encoder, automatic Frequency and Level ranges, selection of increment digit (F,P,N…)
Phase switching by a square signal with a frequency of 1 kHz – adjustable over ± 180° (resolution 1° )
Linearity error < 1% max at 200 kHz – Duty cycle 10-90% for F <1 kHz and 30-70% for F <10 kHz
Output of sweep signal on BNC "Sweep Out" – Level 3.5 V approx.., impedance 10 k approx.
"Sawtooth" or "Triangle" mode – Unlimited excursion between "F Start" & "F Stop" (256 steps)
Rise time < 10 ns (< 7 ns typ.) – Duty cycle 10-90% for F < 200 kHz and 20-80% for F <1 MHz
1 to 65,535 pulses – Synchro/Period by a TTL signal with a frequency < 1 MHz (VCG IN input)
Modulation by signal < 5 kHz, amplitude ± 10 V for modulation from 0 to 100% (VCG IN input)
Modulation by a sine signal with a frequency of 1 kHz – Modulation percentage 20 % or 80 %
TTL & Sweep Out outputs
MIDDLE EAST
Chauvin Arnoux Middle East
P.O. BOX 60-154
1241 2020 JAL EL DIB (BEYROUTH)
Tel : +961 1 890 425
Fax : +961 1 890 424
camie@chauvin-arnoux.com
www.chauvin-arnoux.com
LCD (125 x 45 mm) – Adjustable brightness – Frequency display on 5 digits 20mm high
50 ± 3% / Protection against short-circuits and external voltages up to 60 Vdc or 40 Vac
Sweep by signal < 15 kHz, amplitude ± 10 V – VCF IN input impedance 10 k approx.
Distortion < 0.05 % typically for f < 50kHz, and harmonics < -41 dBc for 50 kHz < f < 1 MHz
Modulation by a signal with a frequency < 15 kHz – Amplitude ± 10 V (VCG IN input)
Switching between "F Start" & "F Stop" by a square signal with a frequency of 1 kHz
50 mVrms for F <30 MHz, 60 mVrms for 30 MHz < F < 80 MHz, 90 mVrms beyond
VCF In input
5-digit display – resolution from 1 mHz to 1 kHz depending on frequency range
230 V ±10 % (or 115 V ±10 %) – 50/60 Hz – 20 VA max. – Removable lead
GX310
Rise time < 10 ns (< 5 ns typ.) – Max. admissible load > 10 TTL loads
Safety as per IEC 61010-1 (2001) – EMC as per EN 61326-1 (2004)
1 to 65,535 pulses – Pulse train period from 10 ms to 100 s
Storage/recall of 15 complete instrument configurations
± 10 Vdc with open circuit – accuracy ± 5% ± 5mV
± 20 ppm for F>10kHz , ± 30 ppm for F<10kHz
"USB A/B" link for the programmable versions
CAT I, 300V / 300 Vrms from 5 Hz to 5 kHz
227 (L) x 130 (H) x 190 (P) – Weight 2.8 kg
Adjustable sweep time from 10 ms to 100 s
State at delivery
1 function generator
1 mains power lead
1 CD-ROM containing:
1 USB A/B lead (programmable versions)
LIN (linear) or LOG (logarithmic)
Operating manual in 5 languages,
FR+GB programming manual,
Labwindows CVI / LabView drivers,
1 M , 22 pF approx.
3 years – France
± 0.05 % + 1 digit
5 Hz to 100 MHz
± 5 ppm / year
± 20 ppm / ° C
0.001 Hz to 20.000 MHz (11 ranges)
VCG, Gate, Clock and Synchro inputs
TTL, Sweep, Clock and Synchro outputs
For information and ordering
GX320