FLUKE-192C/003 Fluke, FLUKE-192C/003 Datasheet - Page 4

no-image

FLUKE-192C/003

Manufacturer Part Number
FLUKE-192C/003
Description
SCOPE METER, 60MHZ, 2 CHANNEL, 500MSPS
Manufacturer
Fluke
Type
Digitalr
Datasheet

Specifications of FLUKE-192C/003

Scope Type
Digital Hand Held
Scope Channels
2
Bandwidth
60MHz
Meter Display Type
LCD Monochrome
Sampling Rate
500MSPS
Input Impedance
1Mohm
Rise Time
5.8ns
Input Voltage
1kV
Brand/series
ScopeMeter®/190 Series
Configuration
Handheld
Display Type
Color
Memory
3 Kpts
Sample Rate
0.5 GS/s
Lead Free Status / RoHS Status
na
Lead Free Status / RoHS Status
na
Memory save and recall
Scope memories: 15 memory locations that each
can contain two waveforms plus corresponding setup.
With each storage action, a user specified name
(20 ASCII characters long) can be assigned to the
stored data, for easier reference.
Recorder memories: 2 memory locations that each
can contain 100 captured dual input scope screens,
or a dual input ScopeRecord (27,500 Min-Max pairs
per input), or a dual input Trendplot (18,000 min-max
pairs + average values)
Real-time clock
Time and date stamp for ScopeRecord, 100 captured
screens and TrendPlots
Mechanical data
Size: 256 mm x 169 mm x 64 mm
(10.1 in x 6.6 in x 2.5 in)
Weight: 2 kg (4.4 lbs)
Case
Design: Rugged, shock proof with integrated
protective holster
Drip and dust proof: IP51 according to IEC529
Shock and vibration: Shock 30g, Vibration
(sinusoidal) 3g according to MIL-PRF-28800F Class 2
Display: Bright full-color LCD with backlight
Brightness: 80 Cd/m
Display size: 115.2 mm x 86.4 mm (4.54 in x 3.4 in);
144 mm (5.67 in) diagonal
Resolution: 320 x 240 pixels
Contrast and brightness: User adjustable,
temperature compensated
Brightness: 80 cd/m
Power
Line power: Country-specific line voltage
adapter/battery charger included
Battery power: Rechargeable NiMH (installed)
Battery operating time: 4 hours
Battery charging time: 4 hours
Battery power saving functions: Auto power down
with adjustable power down time. On-screen battery
status indicator.
Safety
Compliance:
EN61010-1-2001, Pollution degree 2;
UL61010B, with approval;
CAN/CSA C22.2, No. 61010-1-04, with approval;
ANSI/ISA-82.02.01
4 Fluke Corporation
ScopeMeter 215C, 225C and 190C Series, 120 Series
2
2
typ. using power adapter
typ. using power adapter
Input voltage ratings
Maximum probe voltage: CAT II 1000 V/ CAT III 600 V
(Maximum voltage between 10:1 probe tip [VPS200]
and reference lead)
Floating voltage: CAT II 1000 V/ CAT III 600 V
(Maximum voltage between earth ground and any
terminal [signal input or shielding])
Independently isolated inputs:
CAT II 1000 V/ CAT III 600 V
(Maximum voltage between any terminal of one input
or probe [VPS200] and any other terminal of another
input or probe [VPS200])
Maximum voltage on BNC input directly
(input A or B): CAT III 300 V
Maximum voltage on meter input:
CAT II 1,000 V/ CAT III 600 V
Environmental
Operating temperature: 0 °C to +50 °C
Storage temperature: -20 °C to +60 °C
Humidity:
10 °C to 30 °C: 95 % RH non condensing
30 °C to 40 °C: 75 % RH non condensing
40 °C to 50 °C: 45 % RH non condensing
Maximum operating altitude: 3,000 m (10,000 feet)
Maximum storage altitude: 12 km (40,000 feet)
Electro-Magnetic-Compatibility (EMC): EN 61326-1
for emission and immunity
Optically-isolated PC/printer interface
To printer: Supports HP Laserjet,
Epson FX/LQ, Seiko DPU-414 and Postscript printers
via optional PAC 91
To PC: Transfer instrument settings, screen images
and waveform data, compatible with FlukeView
software for Windows
PM9080.
Warranty
Three-years, parts and labor on mainframe instrument
One-year on accessories
Accessories
Standard accessories
Rechargeable battery
pack (installed)
Line voltage adapter/
battery charger
Voltage probes and
accessories
Multimeter test leads
User manual
215C and 225C additional accessories
Three breakout adapt-
ers for bus systems
using RJ45, DB9, and
M12 connectors.
Fluke 215C, 225C and 190C Series
BP190
BC190
10:1 voltage probe (VPS200, 1 red + 1 grey) including hook clip,
ground lead with mini alligator clip, ground lead with hook clip,
4 mm add-on probe tip, ground lead to 4 mm banana plug
TL75 Hard Point test lead set (1red, 1 black)
10 language versions on CD-ROM, “Getting Started” booklet
included with instrument
BHT190
®
via optional OC4USB or
®
DeskJet,
®