N2903A AGILENT TECHNOLOGIES, N2903A Datasheet - Page 20

Removable Hard Disk Drive

N2903A

Manufacturer Part Number
N2903A
Description
Removable Hard Disk Drive
Manufacturer
AGILENT TECHNOLOGIES
Datasheet

Specifications of N2903A

For Use With
Infiniium 9000 Series Oscilloscopes
Lead Free Status / RoHS Status
na
Infi niium 9000 Series performance characteristics
20
Acquisition
Maximum real-time sample rate
Memory depth per channel
Sampling Modes
Filters
Acquisition: digital channels
Maximum real time sample rate
Maximum memory depth per channel
Minimum width glitch detection
Trigger: scope channels
Trigger sources
Sensitivity
Trigger level range – scope channel 1, 2, 3, 4 ± 4 div from center screen (50 Ω and 1 MΩ, all modes except edge)
Sweep modes
Display jitter (displayed trigger jitter)
Trigger holdoff range
Trigger actions
Trigger coupling
1. Internal edge trigger mode. Trigger threshold = fixed voltage at 50% level. The slew rate independent value in the formula represents the traditional trigger jitter.
2. Display jitter example. At 100 mV/div typical noise values are 3.2 mV RMS for 9404 models, 2.5 mV RMS for 9254A models, and 2.4 mV RMS for 9104A models. For slew rate of 500 mVpp
Standard
Option 50M
Option 100
Option 200
Option 500
Real-time
Real-time with peak detect
Real-time with high resolution
Real-time with roll mode (200 ms to 20 sec.)
Equivalent-time (1.0 ps fine interpolator resolution yields a maximum effective sample rate of 1,000 GSa/s)
Segmented memory ( 1 ps time stamp resolution between segments)
Internal
Auxiliary
sin wave with frequency equal to max analog bandwidth of each model, typical display jitter is .95 ps RMS for 9404A models, .97ps for 9254A models, and 1.7 ps RMS for 9104A models.
Up to 8192 segments for 20 Mpts standard memory, up to 131,072 segments with Option 500
Maximum time between triggers is 562,950 seconds (6.5 days)
Re-arm time (minimum time between trigger events) is 4.5 µs with analog channels, 5.8 µs with digital channels on
1 MΩ: DC, AC, (10 Hz) low frequency reject (50 kHZ high pass filter), high frequency reject (50 kHz low pass filter)
1,2
9104, 9254, 9404
4 ch x 10 GS/s or 2 ch x 20 GS/s
20 Mpts on 4 channels, 40 Mpts on 2 channels
50 Mpts on 4 channels, 100 Mpts on 2 channels
100 Mpts on 4 channels, 200 Mpts on 2 channels
200 Mpts on 4 channels, 400 Mpts on 2 channels
500 Mpts/ 250 Mpts on 4 channels, 1 Gpts/ 500 Mpts on 2 channels (single/repetitive mode)
Sin (x) / x Interpolation
2 GSa/s
128/ 64 Mpts with 2 GSa/s. 64/32 Mpts with sampling < 2 GSa/s (single/repetitive mode).
2 ns
Channel 1, channel 2, channel 3, channel 4, aux, and line
1 MΩ input, edge trigger,
50 Ω
Auxiliary
± 8 div from center screen (1 MΩ, edge mode)
± 5 V (50Ω up to 500 MHz with at least 500 mV signal swing)
Auto, triggered, single
100 ns to 10 s fixed and random
Specify an action to occur, and the frequency of the action, when a trigger condition occurs.
Actions include: e-mail on trigger and execute “multipurpose” user settings
DC to 500 MHz: 0.6 div
DC to 2 GHz, 0.5 div
2 GHz to 4 GHz: 1.0 div
DC to 700 MHz: 300 mVp-p
9064
4 ch x 5 GSa/s or 2 ch x 10 GSa/s
9064
9104
9254
9404
.50
.35
.50
.40