U3400A-1CM AGILENT TECHNOLOGIES, U3400A-1CM Datasheet
U3400A-1CM
Specifications of U3400A-1CM
U3400A-1CM Summary of contents
Page 1
Features Up to 120,000 counts resolution Up to 0.012% basic DCV accuracy 11 basic measurements and up to six built-in math functions Dual display on bright VFD Selectable resolutions for variable measurement speeds* Kensington lock slot security * U3402A only ...
Page 2
Take a closer look Input terminals and current fuse Figure 1 Front panel of the U3402A. The front panel of the U3401A is similar with slight differences in the position of certain functions. Please refer to the U3401A’s User’s and ...
Page 3
Typical dual display combinations and applications Primary display Secondary display DCV ACV ACV + DCV DCV DCV DCI DCV ACI ACI + DCI DCV ACV DCI ACI + DCI ACV ACV ACI ACV Hz ACI Hz DCI ACI ACI + ...
Page 4
U3400 Series specifications DC voltage DCV resolution, full-scale reading and accuracy [± reading + count)] U3401A 4.5-digit DMM Range Resolution 500. µV 5.0000 V 100 µV 50.000 500. 1000.0 V ...
Page 5
AC voltage (true RMS, AC coupling mode) ACV resolution, full-scale reading and accuracy [± reading + count)] U3401A 4.5-digit DMM Range Resolution 500. µV 5.0000 V 100 µV 50.000 500. ...
Page 6
AC voltage (true RMS, AC+DC coupling mode) AC+DC voltage resolution, full-scale reading and accuracy [± reading + count)] U3401A 4.5-digit DMM Range Resolution 500. µV 5.0000 V 100 µV 50.000 500. ...
Page 7
DC current DCI resolution, full-scale reading and accuracy [± reading + count)] U3401A 4.5-digit DMM Range Resolution 500.00 µ 5.0000 mA 100 nA 50.000 mA 1 µA 500. µA 5.0000 A 100 µA 10.000 ...
Page 8
AC current (true RMS, AC coupling mode) ACI resolution, full-scale reading and accuracy [± reading + count)] U3401A 4.5-digit DMM Range Resolution Maximum reading 500.00 µ 510.00 5.0000 mA 100 nA 5.1000 50.000 mA 1 µA ...
Page 9
AC current (true RMS, AC+DC coupling mode) AC+DC current resolution, full-scale reading and accuracy [± reading + count)] U3401A 4.5-digit DMM Range Resolution Maximum reading 500.00 µ 510.00 5.0000 mA 100 nA 5.1000 50.000 mA 1 ...
Page 10
Resistance Resistance resolution, full scale reading, and accuracy [± reading + count)] U3401A 4.5-digit DMM Range [1] Resolution 500. 5.0000 kW 100 50.000 500.00 kW 100 W 5.0000 MW ...
Page 11
Diode test/continuity U3401A 4.5-digit DMM Range Resolution Maximum reading 2.3000 V 100 µV 2.3000 V U3402A 5.5-digit DMM Rate Resolution Maximum reading Slow 1.19999 V 10.0000 µV Medium 2.4999 V 100.00 µV Fast 2.499 V 1.000 mV Frequency Frequency resolution, ...
Page 12
Decibel (dB) calculation Range and accuracy (±dB) U3401A 4.5-digit DMM Voltage range [1][2] Input voltage 500. 500 mV 5.0000 V 5000 50.000 500. ...
Page 13
Supplemental specifications Full-scale display counts Model Reading rate U3401A N/A U3402A Slow Medium Fast Supplemental measurement specifications Measurement DC voltage Measurement method Input resistance Maximum input voltage Input protection Response time DC current Shunt resistance Maximum input and overload protection ...
Page 14
Supplemental measurement specifications Measurement AC+DC Measurement method voltage Crest factor Input impedance Maximum input voltage Overload ranging Input protection Response time AC current Measurement method Crest factor Shunt resistance Input protection Response time AC+DC Measurement method current Crest factor Measurement ...
Page 15
Supplemental measurement specifications Measurement Resistance Input protection Response time Diode/ Measurement method Continuity Test current Open-circuit voltage Continuity threshold Continuity level Audible tone Input protection Resistance/ Measurement method Continuity Test current (U3402A Open-circuit voltage only) Audible tone Zeroing error Input ...
Page 16
Supplemental measurement specifications Measurement dBm 0 dBm operation Resolution Reference impedance [1] Math operation I/O interface [1] Reference impedance is displayed on the secondary display. [2] Reading is displayed in watts (audio power). Specification 600 W reference ...
Page 17
... RS-232 (for calibration use only) Dimensions (W × H × D) 255 mm × 105 mm × 305 mm (with bumpers) 215 282 mm (without bumpers) Weight 3.44 kg (with bumpers) Warranty One year for U3401A/U3402A Three months for standard shipped accessories Calibration cycle One year Warm-up time At least 30 minutes ...
Page 18
... Quick Start Guide Product Reference CD Certificate of Calibration 34138A Test Lead Kit AC Power Cord Accessories 34138A Test Lead Kit 34133A Precision Electronic Test Leads U1161A Extended Test Lead Kit 11059A Kelvin Probe Set 18 34330A Current Shunt (30 A) U3400A-1CM Rack Mount Kit ...
Page 19
... Product specifications and descriptions in this document subject to change without notice. www.agilent.com www.agilent.com/find/low- cost-dmm For more information on Agilent Technologies’ products, applications or services, please contact your local Agilent office. The com- plete list is available at: www.agilent.com/find/contactus Phone or Fax Americas ...