Performance characteristics
* Denotes warranted specifications, all others are typical.
Horizontal system digital channels
Characteristics
Maximum sample rate
Maximum record length
Minimum detectable pulse width
Channel-to-channel skew
Trigger system
Characteristics
Trigger modes
Trigger coupling
Trigger holdoff range
Trigger level range
Specifications are valid after a 30-minute warm-up period and ±10 °C from firmware calibration temperature.
Trigger sensitivity
Internal*
External*
Any channel
External
All MSO models and DSO models with MSO upgrade
1 GSa/s
2 Mpts (digital only), 500 kpts (analog and digital channels)
5 ns
2 ns (typical); 3 ns (maximum)
All 3000 X-Series models
• Normal (triggered): requires trigger event for scope to trigger
• Auto: triggers automatically in absence of trigger event
• Single: triggers only once on a trigger event, press [Single] again for scope to find another
• Force: front panel button that forces a trigger
DC: DC coupled trigger
AC: AC coupled trigger, cutoff frequency: < 10 Hz (internal); <50 Hz (external)
HF Reject: High frequency reject, cutoff frequency ~ 50 kHz
LF Reject: Low frequency reject, cutoff frequency ~ 50 kHz
Noise Reject: Selectable OFF or ON, decreases sensitivity 2x
40 ns to 10.00 s
<10 mV/div: greater of 1 div or 5 mV; ≥ 10 mV/div: 0.6 div
200 mVpp from DC to 100 MHz
350 mVpp 100 MHz to 200 MHz
± 6 div from center screen
± 8 V
trigger event, or press [Run] to trigger continuously in either Auto or Normal mode