N6708A AGILENT TECHNOLOGIES, N6708A Datasheet - Page 20

Filler Panel Kit

N6708A

Manufacturer Part Number
N6708A
Description
Filler Panel Kit
Manufacturer
AGILENT TECHNOLOGIES
Datasheet

Specifications of N6708A

Required When Using Less Than 4 Modules
Accessory Type
For Use With
Agilent Type 6700B Series DC Module Power Supplies
Lead Free Status / RoHS Status
na
Maximum total
output power
(= sum of total
module output
power)
Command
processing time
Protection response
characteristics
Digital control
characteristics
Interface capabilities
Environmental conditions
Agilent N6700B, N6701A, N6702A MPS Mainframes
N6700B, N6701A, N6702A
N6700B
N6701A
N6702A
From receipt of command to start
of the output change
INH input
Fault on coupled outputs
Maximum voltage ratings
Pins 1 and 2 as FLT output
Pins 1 - 7 as digital/trigger outputs
(pin 8 = common)
Pins 1 - 7 as digital/trigger inputs and
pin 3 as INH input
(pin 8 = common)
GPIB:
LXI compliance
USB 2.0
10/100 LAN
Built-in web server
Operating environment
Temperature range
Relative humidity
Altitude
Storage temperature
LED statement
1200 W
≤ 1 ms
< 10
16.5 VDC/- 5 VDC between pins (pin 8 is internally connected to chassis ground).
Maximum low-level output voltage = 0.5 V @ 4 mA
Maximum low-level sink current = 4 mA
Typical high-level leakage current = 0.14 mA @ 16.5 VDC
Maximum low-level output voltage = 0.5 V @ 4 mA; 1 V @ 50 mA; 1.75 V @ 100 mA
Maximum low-level sink current = 100 mA
Typical high-level leakage current = 0.12 mA @ 16.5 VDC
Maximum low-level input voltage = 0.8 V
Minimum high-level input voltage = 2 V
Typical low-level current = 2 mA @ 0 V (internal 2.2 k pull-up)
Typical high-level leakage current = 0.12 mA @ 16.5 VDC
SCPI - 1993, IEEE 488.2 compliant interface
Class C (applies to mainframes with firmware revision C.00.02 and up)
Requires Agilent IO Library version M.01.01 and up, or 14.0 and up
Requires Agilent IO Library version L.01.01 and up, or 14.0 and up
Requires Internet Explorer 5+ or Netscape 6.2+
Indoor use, installation category II (for AC input), pollution degree 2
0°C to 55°C (current is derated 1% per °C above 40°C ambient temperature)
Up to 95%
Up to 2000 meters
-30°C to 70°C
Any LEDs used in this product are Class 1 LEDs as per IEC 825-1
5 µs from receipt of inhibit to start of shutdown
400 W
600 W
600 W
µ
s (from receipt of fault to start of shutdown)
20
when operating from 100 – 240 VAC input
when operating from 100 – 240 VAC input
when operating from 200 – 240 VAC input
when operating from 100 – 120 VAC input