CPX400DP AIM-TTI INSTRUMENTS, CPX400DP Datasheet - Page 4

no-image

CPX400DP

Manufacturer Part Number
CPX400DP
Description
PSU DUAL 60V MAX 20A MAX PROG
Manufacturer
AIM-TTI INSTRUMENTS
Datasheet

Specifications of CPX400DP

Power Supply Output Type
Programmable
No. Of Output Channels
2
Output Voltage
60V
Output Voltage 2
60V
Output Current
20A
Output Current 2
20A
Power Rating
840W
Input Voltage
115VAC / 240VAC
No. Of Outputs
2
Rohs Compliant
NA
Lead Free Status / RoHS Status
na
Specifications CPX400D & CPX400DP
Voltage/Current/Power Levels
Voltage Range:
Current Range:
Note: Actual maxima for voltage and current are typically 1% greater than the figures given above.
Power Range:
POWER ENVELOPE
The maximum current atany voltage settings is limited by the power envelope which is set to
give 7A at 60V rising to 20A at 20V under all ac supply conditions (both outputs loaded). At
lower output voltages the power is restricted by the 20 amps current maximum.
See PowerFlex power envelope graph on previous page.
Output Setting & Control
Voltage Setting:
Current Setting:
Output Mode:
Output Switch:
View Settings:
Status Indication:
S-Lock
(Settings Lock) Voltage and current settings can be locked by a single button press.
Lock accuracy is equal to the meter accuracy (see Meter Specifications).
Output Performance
Ripple & Noise:
Load Regulation:
Line Regulation:
Transient Response: <250µs to within 50mV of setting for a 5% to 95% load change.
Temp. Coefficient:
Output Protection
Output Protection:
OVP Setting/Range: Via screwdriver adjustable preset on front panel. Range 1V to 66V
Over-temperature:
Safety Interlocks:
Output Connections
Output Terminals:
Terminals can accept fixed shroud 4mm plugs, standard 4mm plugs, fork terminals and bare wires.
Remote Sense
Sense Selection:
Sense Terminals:
Display Type:
Voltage Meter
Resolution/ Accuracy: 10mV / ± 0·1% of reading ± 2 digits
Current Meter
Resolution/ Accuracy: 10mA / ± 0·3% of reading ± 20mA
Independent Mode
In the normal mode of operation, each output is fully independent and isolated. Operation
is equivalent to two single output power supplies.
Voltage Tracking Mode
The two outputs remain isolated, but the Slave voltage controls are disabled and the Slave
voltage is set equal to the Master voltage. This can be used to generate tracking bipolar
voltages, or tracking unipolar voltages relative to different grounds.
When voltages greater than 60V are required, the outputs can be wired in series to generate
0 to 120V with the voltage controlled from the Master.
When currents greater than 20A are required, the outputs can be wired in parallel to create
the equivalent of a 40A power supply with the voltage controlled from the Master.
Track Accuracy:
OUTPUT SPECIFICATIONS (each output)
METER SPECIFICATIONS (each output)
VOLTAGE TRACKING
Designed and built in Europe by:
Glebe Road, Huntingdon, Cambridgeshire. PE29 7DR United Kingdom
Email: sales@tti-test.com Web: www.tti-test.com
Tel: +44 1480 412451 Fax: +44 1480 450409
Thurlby Thandar Instruments Ltd.
0V to 60V.
0A to 20A.
Up to 420W subject to power envelope.
By coarse and fine controls.
By single logarithmic control.
Constant voltage or constant current with automatic cross-over.
CC indicator lit in constant current mode.
Electronic, non isolating. Preset voltage and current limit displayed
when Output is off. Output rise time no load <10ms.
With the output On, the meters show actual voltage and current.
The preset levels can be viewed and adjusted at any time by pressing
the View Settings button.
LED indication of Output On, V/I Limits, CV, CI, Power Limit, Remote,
LAN status. Message on meter display for trip condition.
Typically <3mV rms, <15mV pk-pk, (5mV rms max.) in CV mode.
Voltage - <0·01% of maximum output for any load change within
the PowerFlex envelope (remote sense connected).
Current - <0·05% of maximum output for any load change within
the PowerFlex envelope.
Voltage - <0·01% of maximum output for a 10% line change.
Current - <0·01% of maximum output for a 10% line change.
Typically <100ppm/°C
Forward protection by Over-voltage Protection (OVP) trip.
Reverse protection by diode clamp for currents to 3A.
Output trips off for over-temperature.
Operations that could cause an unexpected change in voltage or
current settings are interlocked with the output switch.
Universal 4mm safety binding posts on 19mm (0·75”) spacing at front.
Screw terminals at rear (CPX400DP only).
Voltage sensing is selected as Local or Remote by front panel switch.
Sprung loaded screw-less terminals at front.
Screw terminals at rear (CPX400DP only).
Dual 4-digit meters, 10mm (0·39”) LED.
Slave voltage = ± (0·1% of Master voltage setting + 10mV)
Each output has an independent DC On/Off control, however, an additional control button is
provided which turn both outputs on or off simultaneously.
The CPX400DP offers full remote control and read-back using RS-232, USB, GPIB or LAN
(compliant with LXI class C). All interfaces are at ground potential and opto-isolated from the
output terminals. Note: Remote/Local Sense, is manually selectable only.
RS-232
Standard 9-pin D connector. Baud rate 9,600.
USB
USB 2.0 connection (backwards compatible with USB 1.x). Operates as a virtual COM port.
GPIB (IEEE-488)
The interface conforms with IEEE-488.1 and IEEE-488.2.
Ethernet (LAN)
Standard 10/100 base-T hardware connection. ICMP and TCP/IP Protocol for connection to
Local Area Network or direct connection to a single PC.
LXI Compliance
LAN interface is compliant with LXI class C. (LXI is the abbreviation for Lan eXtensions for
Instrumentation). For more information visit: www.tti-test.com/go/lxi
Voltage Setting
Resolution/Accuracy: 1mV / ± (0.05% +10mV)
Current Setting
Resolution/Accuracy: 1mA / ± (0.3% +5mA)
Programming Speed
Command Delay:
Voltage Up Time:
Voltage Down Time: <80ms* to 1% (full load); <1.5s* to 1% (no load)
* The up and down times vary with range and voltage step size. More information is
contained in the operating manual which can be downloaded from our web site.
IVI Driver
An IVI driver for Windows is supplied. This provides support for common applications such as
LabView*, LabWindows*, HPVEE* etc.
USB Driver
An installation file is supplied which calls a standard Windows* USB driver.
* LabView and LabWindows are trademarks of National Instruments.
* USB interface is supported for Windows 2000, XP, and Vista.
Input
AC Input:
Input Power:
Temperature & Environmental
Operating Range:
Storage Range:
Environmental:
Cooling:
Safety & EMC
Safety:
EMC:
Physical
Size:
Weight:
Rack Mount (RM300A)
19 inch 3U rack mount suitable for one or two power supplies.
Accuracy specifications apply for the temperature range 18
Thurlby Thandar Instruments Ltd. operate s a policy of continuous development and reserves the
right to alter specifications without prior notice.
BOTH ON / BOTH OFF
DIGITAL BUS INTERFACES (CPX400DP only)
DIGITAL PROGRAMMING PERFORMANCE (CPX400DP only)
DRIVER SOFTWARE SUPPLIED (CPX400DP only)
HPVEE (now Agilent VEE) is a trademark of Agilent Technologies.
GENERAL SPECIFICATIONS
OPTIONS
Windows is a trademark of Microsoft.
Typically <25ms (this must be added to any of the figures below)
<10ms* to 1%
110 to 240 volts ±10% 50/60Hz. Installation Category II.
1250VA max.
+5ºC to +40ºC, 20% to 80% RH
-40ºC to + 70ºC
Indoor use at altitudes up to 2000m, Pollution Degree 2.
Rear discharge variable speed fan.
Complies with EN61010-1
Complies with EN61326
210 x 130 mm (½ rack 3U) x 375mm
(size excludes feet, knobs and terminals).
6.1kg (CPX400P); 6.3kg (CPX400DP)
o
C to 28
Brochure Part No. 82100-XXXX
o
C after one hour warm-up.