QPX1200L AIM-TTI INSTRUMENTS, QPX1200L Datasheet - Page 2

1200W PowerFlex Max 60V Or 50A

QPX1200L

Manufacturer Part Number
QPX1200L
Description
1200W PowerFlex Max 60V Or 50A
Manufacturer
AIM-TTI INSTRUMENTS
Datasheet

Specifications of QPX1200L

Power Supply Output Type
Variable
No. Of Outputs
1
Output Voltage
60V
Output Current
50A
Power Rating
1.2kW
Input Voltage
110V AC To 240V AC
Load Regulation
0.01%
Length
413mm
Width
350mm
QPX1200L 1.2kW PowerFlex dc power supply - 60V, 50A max.
Unrivalled Flexibility
The QPX1200L offers users a level of flexibility that can not be
achieved with conventional laboratory power supplies.
With a current capability of 20 amps at the maximum output of
60 volts, its PowerFlex design offers increasing output current
with reducing output voltage. The QPX1200L can therefore per-
form the task of many different power supplies.
Example voltage/current combinations include 60V/20A,
48V/25A, 37.5V/30A, 26V/40A, and 20V/50A.
High precision
With setting controlled by an instrumentation quality 16 bit DAC,
the QPX1200L offers high accuracy and stability.
Low noise
Mixed mode regulation with a linear output stage gives the
QPX1200L exceptionally low output noise for a unit of this
power, and ensures good transient response.
Bench or rack mounting
Unlike most higher power laboratory PSUs, the QPX1200L is
equally suited to bench or rack-mount applications.
Output and sense terminals are provided on both the front and
rear panels. The case has integral tilt feet, and a 3U rack mount
is available as an option.
A wide range of interfaces
The QPX1200 has analog, RS-232, USB and LAN interfaces as
standard, the latter conforming with the LXI standard.
An optional GPIB interface can be specified at time of order.
Voltage Range:
Current Range:
Power Range:
Operating Mode:
Voltage Setting:
Current Setting:
Setting Accuracy:
Setting Stores:
Load regulation:
Line regulation:
Ripple & Noise:
Transient Response: <250µs to within 50mV of set level for 90% load change.
Temp. Coefficient:
Output Protection:
Protection Functions: Power Limiter, over-voltage, over-current and temperature
Status Indication:
Output Switch:
Output Terminals:
Sensing:
OUTPUT SPECIFICATIONS
Designed and manufactured in Europe by:
Tel: +44 (0)1480 412451 Fax: +44 (0)1480 450409
Email: sales@tti-test.com Web: www.tti-test.com
Glebe Road, Huntingdon. Cambs. PE29 7DR U.K.
Thurlby Thandar Instruments Ltd.
0V to 60V.
0A to 50A.
Up to 1200W - see PowerFlex power envelope graph.
Constant voltage or constant current with automatic
cross-over and mode indication.
By floating point numeric entry or rotary jog wheel;
resolution 1mV.
By floating point numeric entry or rotary jog wheel;
resolution 10mA.
Voltage - 0.1% ± 2mV. Current 0.3% ± 20mA
Up to 10 set-ups can be saved and recalled via the keyboard
or remote interfaces.
<0.01% +5mV for a 90% load change.
<0.01% +5mV for a 10% line voltage change.
Typically <3mV rms, <20mV pk-pk, (20MHz bandwidth).
Typically <100ppm/
Forward protection by OVP trip; maximum voltage that
should be applied to the terminals is 70V. Reverse protection
by diode clamp.
trips.
Display indication of Output On, CV, CI and Power Limit.
Message on display for over-voltage trip.
Push switch operating electronic power control.
LED indication of On state
Front and rear output and sense terminals. Safety terminals
accepting 6mm wire diameter, 6mm plugs or 8mm spades at
50 Amps max, or 4mm plugs at 30 Amps max.
Selectable local or remote sensing.
o
C.
Output Protection:
Fault Condition Trip:
Over Voltage (OVP): Settable 2V to 65V in 0.1V steps
Over Current (OCP): Settable 2A to 55A in 0.1A steps
Over Temperature:
Sense Error:
Display Type:
Meter Function:
Limits Display:
Meter Resolution:
Meter Accuracy:
Quasi-Analog:
USB:
RS-232:
LAN:
GPIB (Option G):
N.B. All bus interfaces incorporate full control, readback and status reporting.
Bus Type Selection:
Address Selection:
Baud Selection:
Setting Resolution:
Readback Resolution: Voltage - 10mV, Current 10mA.
Accuracy:
AC Input:
Power Consumption: 1600VA max.
Operating Range:
Storage Range:
Environmental:
Safety:
EMC:
Size:
Weight:
Benchtop Operation: Folding legs are incorporated that can be used to angle the
Rack Mount Option:
The maximum current at
any voltage settings is
limited by the power
envelope which is set to
give 20A at 60V rising to
50A at 20V under all ac
supply conditions.
At lower output voltages
the power is restricted by
the 50 amps current
maximum.
Thurlby Thandar Instruments Ltd. operates a policy of continuous development and reserves
the right to alter specifications without prior notice.
OUTPUT PROTECTION
METERING
BUS INTERFACES
GENERAL
POWER ENVELOPE
Conforming to Ethernet standard and LXI standard.
Output will withstand forward voltages of up to 70V a. Reverse
protection by diode clamp.
The output will be shut down if any of the four trip conditions
listed below occur
Monitors internal temperature rise to protect against excess am-
bient temperature or blocked ventilation slots.
Monitors the voltage between the remote sense terminals and
output terminals to protect against mis-wiring.
Large dot-matrix black-on-white backlit LCD.
5 digit voltage meter and 4 digit current meter. Display of
limits values, memory contents etc.
Wide range of alpha-numeric messages.
Limits settings can be displayed simultaneously with mea-
surements.
Voltage: 1mV. Current: 10mA
Voltage: 0.1% of reading ± 2 digits
Current: 0.3% of reading ± 2 digits
Voltage or current, 0 - 5V or 0 - 10V. (Digitised)
Standard USB harware connection.
Variable baud rate 19,200 max.
Conforming with IEEE-488.1 and IEEE-488.2.
From front panel (RS232/USB/LAN/GPIB).
From front panel (1 to 31).
RS-232 only. From front panel (600 to 19200 baud).
Voltage - 1mV, Current 1mA.
See specifications under Outputs and Metering.
110V to 240V AC ±10%, 50/60Hz. Installation Category II.
+5
-40
Indoor use at altitudes to 2000m, Pollution Degree 2.
Complies with EN61010-1.
Complies with EN61326.
350 x 130 x 413mm (WxHxD) (3U height) .
9.2kg (20lb).
front panel upwards when required.
19 inch 3U rack mount.
o
C to +40
o
C to + 70
60V
50V
40V
30V
20V
10V
0V
0A
o
C, 20% to 80% RH.
o
C.
5A
10A
15A
20A
25A
QPX1200
30A
PowerFlex
Power Envelope
82100-1320 Iss. 2
35A
(1200W max.)
40A
45A
50A

Related parts for QPX1200L