N3305A AGILENT TECHNOLOGIES, N3305A Datasheet

DC ELECTRONIC LOAD MODULES, 500W

N3305A

Manufacturer Part Number
N3305A
Description
DC ELECTRONIC LOAD MODULES, 500W
Manufacturer
AGILENT TECHNOLOGIES
Datasheet

Specifications of N3305A

Power Supply Output Type
Programmable
Power Rating
500W
Input Voltage
0 V To 150 V
Weight
10lb
Calibrated
No
Input Current Min
0A
Input Current Max
60A
Supply Voltage
150V
Power Supply Mounting
Rack Mount
Rohs Compliant
NA
Lead Free Status / RoHS Status
na
Increase your Manufacturing Test Throughput with
Fast Electronic Loads
Agilent dc Electronic Loads
Models N3300A-N3307A
Technical Specifications
• Increase test system throughput
• Lower cost of ownership
• Decrease system development time
• Increase system reliability
• Increase system flexibility
• Stable operation down to zero volts
• dc connection terminal for ATE applications
Agilent Technologies

Related parts for N3305A

N3305A Summary of contents

Page 1

... Increase your Manufacturing Test Throughput with Fast Electronic Loads • Increase test system throughput • Lower cost of ownership • Decrease system development time • Increase system reliability • Increase system flexibility • Stable operation down to zero volts • dc connection terminal for ATE applications Agilent Technologies ...

Page 2

Increase Test Throughput Reduced command processing time: Automatically execute stored command sequences: Programmable delay allows for either simultaneous or sequential load changes: Buffer measurement data: Control measurement speed vs. accuracy: Control rising and falling slew rates separately: Increase System Flexibility…for ...

Page 3

Test high current power supplies: Control the electronic load how you want to: Quickly create powerful and consistent software: Plug-n-Play Make Measurements Easily and Accurately Measure with all load modules simultaneously: 3 Measure voltage and current simul- taneously: Observe transient ...

Page 4

... 250 W 300 W 500 W 600 W Derated Current Detail Voltage full scale all specifiations apply 3 slew rate limitations apply 2 (see Table A- full scale 120 N3303A N3304A N3305A N3306A 1.2 V 1.2 V 1 N3307A 150 V 250 W Current ˚ C, N3307A 1.4 V ...

Page 5

... Specification is ±(% of reading + fixed offset). Measurement is 1000 samples. Specification may degrade when the unit is subject field of 3 V/meter, the unit is subject to line spikes of 500 electrostatic discharge current accuracy specifications apply 30 seconds after input current is applied. 5 N3304A N3305A N3306A N3307A A/120 A 3 A/30 A ...

Page 6

... So even though the transient mode is still operational at lower voltages, a fast pulse train with large transitions may not be achievable. N3303A N3304A N3305A 0.02 mA/0.2 mA 0.1 mA/1 mA 0.1 mA/1 mA ...

Page 7

... External Analog Programming Voltage Programming 0. Accuracy Current Programming 0.25 Accuracy External Monitor Ports Voltage Monitor 0.25 Accuracy Current Monitor 0.1% + 4.5 mA Accuracy 2 Applies to all ranges. N3303A N3304A N3305A (with specified 20 ms (with specified 20 ms (with specified 10 ms (with < (with < (with < ...

Page 8

... Net Weight: N3300A: 13.2 kg (29 lb); N3301A: 7.3 kg (16 lb); N3302A, N3303A, N3304A or N3307A: 2 lb); N3305A or N3306A: 4.6 kg (10 lb) Shipping Weight: N3300A (38 lb); N3301A: 9.1 kg (20 lb) N3302A, N3303A, N3304A or N3307A: 4 lb) N3305A or N3306A: 6.8 kg (15 lb) 8 N3301A ˚ ˚ 100 - 250 Vac 2 100 - 250 Vac 230 115 Vac ...

Page 9

Agilent N3300A 190.5mm 7.5" Rear 12.7mm 0.5" Agilent N3301A 190.5mm 7.5" Rear 12.7mm 0.5" Top 624.8mm 24.6" Side 50.8mm 2.0" Top 624.8mm 24.6" Side 50.8mm 2.0" 9 425.5mm 16.75" 177.8mm 7.0" 213.4mm 8.4" 177.8mm 7.0" ...

Page 10

... Agilent Technologies’ Test and Measurement Support, Services, and Assistance Agilent Technologies aims to maximize the value you receive, while minimizing your risk and problems. We strive to ensure that you get the test and measurement capabilities you paid for and obtain the support you need. Our extensive support resources and services can help you choose the right Agilent products for your applications and apply them successfully ...