MCR-S10-50-UI-DCI-NC Phoenix Contact, MCR-S10-50-UI-DCI-NC Datasheet - Page 15

no-image

MCR-S10-50-UI-DCI-NC

Manufacturer Part Number
MCR-S10-50-UI-DCI-NC
Description
INTERFACE ANALOG
Manufacturer
Phoenix Contact
Datasheet

Specifications of MCR-S10-50-UI-DCI-NC

Input Type
0 - 9.5A, 0 - 55A
Output Type
Current Or Voltage
No. Of Output Channels
2
Accuracy
± 0.5%
Supply Voltage Range
20VDC To 30VDC
Approval Bodies
CE, UL Recognised
External Depth
114.5mm
External
RoHS Compliant
External Length / Height
99mm
Rohs Compliant
Yes
Fine Adjustment of the Measuring Transducer
When the green "POWER" LED is lit, the operating voltage
of 20 V DC … 30 V DC is connected.
100241_07_en
Signal Input
Close the module after roughly setting the input current
range and preselecting the output signal.
Connect the module to the signal lines and the
operating voltage according to the instructions in
Section "Electrical Connection" on page 10.
Range
10 A
1 A
5 A
When using the MCR-S-…-DCI, ensure you
use the correct signal input terminal block
for your measuring range.
Allow the module to warm up for 2 minutes
before starting the adjustment procedure.
Terminal Block
Signal Input
7
6
5
Ground Terminal
Block
8
8
8
Analog Output
Figure 9
ZERO and SPAN behavior (±25% each) are shown in
Figure 10 and Figure 11:
Figure 10
Figure 11
ZERO potentiometer for zero point adjustment
SPAN potentiometer for final value adjustment
-100
Potentiometer
Offset (ZERO) behavior
Final value (SPAN) behavior
-100
SPAN -25%
OUT
OUT
[%]
110
100
[%]
110
100
-100
-110
M C
A r
-100
-110
SPAN +25%
t . - N
R -
r . :
S -
1 - 5
2 8
- U
1 4
I - S
6 5
W -
D C
0
+25%
PHOENIX CONTACT
I
100
100
SPAN
MCR-S-…-DCI
[%]
-25%
IN
[%]
IN
15

Related parts for MCR-S10-50-UI-DCI-NC