DPM 942 LASCAR, DPM 942 Datasheet

Loop Powered Meter

DPM 942

Manufacturer Part Number
DPM 942
Description
Loop Powered Meter
Manufacturer
LASCAR
Datasheet

Specifications of DPM 942

No. Of Digits / Alpha
3-1/2
Meter Function
4-20mA Loop Current
Meter Range
4mA To 20mA
Digit Height
19mm
Operating Temperature Range
0°C To +50°C
Meter Signal Input
Current
Character Size
0.75"
Height
1.73"
Accuracy
+/- (0.1 Of Input +/- 1 Count)
Rohs Compliant
NA
Lead Free Status / RoHS Status
na
DPM 942
A loop powered indicator giving an accurate, zero corrected indication of 4-20mA loop current. Calibration is by two 20-turn cermet
potentiometers which allow sensitive adjustment of the instrument. The meter is housed in a robust carrier, which can be bolted in
place or panel mounted using the low profile bezel and clips provided.
CUSTOM SCALING
For alternative scaling to the standard 0 - 1000, cut Link 8
and short Link 7. Add two resistors R12 and R13 which
will set the basic scale of the meter. The highest practical
value for the R12 and R13 series combination is 12R. To
select your desired scaling, use the following formula:
R12 + R13 = Full Scale reading at 20mA
Re-calibrate as necessary.
Specifications liable to change without prior warning
DIMENSIONS
19mm (0.75“) Digit Height
Programmable Decimal Points
Low Volt Drop
Simple Screw Terminal Connections
Top right hand corner of the pcb.
Top right hand corner of the pcb.
R25
R25
R14
R14
R13
R13
R12
R12
R11
R11
160
All dimensions in mm (inches)
51.0 (2.01)
76.0 (2.99)
61.5 (2.42)
72.5 (2.85)
51.0 (2.01)
71.5 (2.81)
(0.12)
(0.22)
Ø3.0
Ø5.7
DPM 942
a
b c
CONNECTOR SOURCING GUIDE
d
Standard Meter
Specification
Accuracy (overall error) *
Linearity
Sample rate
Operating temperature range
Temperature s
Loop Volt Drop
*To ensure maximum accuracy, re-calibrate periodically.
METHOD
e
Issue 7
tability
a.
b.
c.
d. 11.5 (0.45)
e. 11.0 (0.43)*
1.0 (0.04)
2.0 (0.08)
6.5 (0.26)
November/2006
Screw Terminals - No Connector Required
OFFSET SPAN
*typical depth of tallest component behind PCB,
this dimension is for guidance only.
Min.
0.05
4-20mA Loop Powered Indicator
3.9
10
6
0
ON BOARD
SOLDER LINKS
1
Panel cut-out 72 x 40 (2.83 x 1.57)
Maximum panel thickness 3mm (0.12).
Typ.
100
4.3
M.C.
REAR
VIEW
5
3
7 8 2
9
Max.
0.1
±1
4.8
50
4
3
Applies to DPM 942/3
No.:695 - Page 1 of 2
Stock Number
%(±1 count)
samples/sec
ppm/°C
DPM 942
count
Unit
°C
V

Related parts for DPM 942

DPM 942 Summary of contents

Page 1

... Panel cut-out (2.83 x 1.57) Maximum panel thickness 3mm (0.12). *typical depth of tallest component behind PCB, this dimension is for guidance only. OFFSET SPAN REAR VIEW BOARD SOLDER LINKS November/2006 M.C. Applies to DPM 942/3 Stock Number DPM 942 Unit %(±1 count) count samples/sec °C ppm/°C V ...

Page 2

... ON-BOARD LINKS: In order to quickly and easily change operating modes for different applications, the meter has several "on-board links". They are designed to be easily opened (cut) or shorted (soldered). The DPM 942 is powered from the 4-20mA signal loop and needs no other power supply. Ensure correct polarity when connecting. PANEL FITTING Specifications liable to change without prior warning LASCAR ELECTRONICS INC ...

Related keywords