T6613-F GE Sensing, T6613-F Datasheet - Page 2

no-image

T6613-F

Manufacturer Part Number
T6613-F
Description
SENSOR, CO2, FLOW, 400-2000PPM
Manufacturer
GE Sensing
Datasheet

Specifications of T6613-F

Supply Voltage Range Dc
4.75V To 5.25V
Accuracy
± 90ppm
Flow Measuring Range
400ppm To 2000ppm
Svhc
No SVHC (18-Jun-2010)
Lead Free Status / RoHS Status
Lead free / RoHS Compliant
Telaire 6613 CO
Specifi cations
Method
Non Dispersive Infrared (NDIR), gold plated optics,
diff usion or fl ow through sampling (with Telaire’s
Patented ABC Logic Self Calibrated Algorithm)
Measurement Range
400 to 2000 ppm
Dimensions
2.25 in X 1.365 in X 0.60 in
(57.15 mm X 34.67 mm X 15.24 mm)
Accuracy*
@72°F (22°C) when compared against a certifi ed factory
reference ±40 ppm + 2% of reading up to 1250 ppm
Temperature Dependence
0.2% FS per °C
Stability
< 2% of FS over life of sensor (15 year typical)
Pressure Dependence
0.135 of reading per mm Hg
Calibration Interval
Not required
Response Time
< 2 minutes for 90% step change typical
Signal Update
Every 4 seconds
Warm Up Time
• < 2 minutes (operational)
• 10 minutes (maximum accuracy)
Operating Conditions
• 32°F to 122°F (0°C to 50°C)
• 0 to 95% RH, non condensing
Storage Conditions
-40°F to 158°F (-40°C to 70°C)
2
Module
Output
Digital
UART @ 19200 Baud (Please call for detailed product
specifi cations).
Analog
0.8 to 4 VDC
Power Supply Requirements
5 VDC regulated (±5%)
Power Consumption
• 0.90 watts peak
• 0.165 watts average
Interface Connections
Designed for 12 pin male header with 0.1 in (2.54 mm)
spacing. Header not included.
Flow Rates (Via Flow Ports)
Flow through version 40 to 50 cc/min
Warranty Terms
12 months
* Handling and OEM assembly may aff ect factory
(9.65 mm)
0.38 in
calibration. Specifi ed accuracy is after 21 days of
continuous operation with ABC Logic.
2 Mounting Holes Ø.105 (2.667)
(57.15 mm)
2.25 in
(44.45 mm)
1.75 in
(28.575 mm)
1.125 in
0.15 in (3.81 mm)
(0.1575 mm)
0.062 in
(15.24 mm)
(13.7 mm)
0.60 in
0.54 in

Related parts for T6613-F