DP120 Fluke Electronics, DP120 Datasheet - Page 2

PROBE VOLTAGE DIFFERENTIAL

DP120

Manufacturer Part Number
DP120
Description
PROBE VOLTAGE DIFFERENTIAL
Manufacturer
Fluke Electronics
Type
Voltage Probe, Differentialr
Datasheet

Specifications of DP120

Includes
Alligator Clips, Battery, Pin-Grabber Test Clips
Attenuation Value
20:1, 200:1
Bandwidth
20MHz
Resistance - Input
1M
Voltage - Max
CAT III 600V, CAT II 1000V
Capacitance - Input
50pF
Cable Length
60.000" (1524.00mm)
Color
Gray
Lead Free Status / RoHS Status
Contains lead / RoHS non-compliant
Other names
614-1104

Available stocks

Company
Part Number
Manufacturer
Quantity
Price
Part Number:
DP120
Manufacturer:
TI/德州仪器
Quantity:
20 000
Company:
Part Number:
DP1200A
Quantity:
3 500
Part Number:
DP1205C433LF
Manufacturer:
MOLEX
Quantity:
5 000
Installing or Replacing the Battery
Replace the battery when the red LED lights, or when both red and
green LEDs are off when the probe is turned on.
Follow the procedure (steps 1 to 5) to install or replace the battery.
To avoid electrical shock, disconnect the differential
voltage probe from any conductor, and disconnect
the ScopeMeter
replacing the battery.
®
test tool or oscilloscope before
Warning
Characteristics
FLUKE guarantees the properties expressed in numerical values
with the stated tolerance. Specified non-tolerance numerical values
indicate those that could be nominally expected from the mean of a
range of identical differential voltage probes.
Input Characteristics
Input Probe Tip Style: Shrouded banana probe tip
Probe Cable Length: 1.5 meter (60 inch)
Installation Category III refers to distribution level and fixed
installation circuits inside a building. Installation Category II refers
to local level, which is applicable for appliances, and portable
equipment.
Output Characteristics
Output Cable: Safety designed BNC cable
Cable Length: 0.5 meter (20 inch)
Max. Output Voltage Range: 6.5V into 1 M
Electrical Characteristics
Attenuation: 200x and 20x
Bandwidth: (into 1 M
Accuracy: 2.5% into 1 M
Rise time:
CMRR:
600V CAT III
1000V CAT II
1000 VDC, or
1000 Vrms, or
1200V (DC+AC peak)
For derating of each input probe (red or black), see figure below.
200x: DC to 20 MHz (-3 dB)
200x: 17.5 ns
200x: @60Hz= >80 dB, @1 MHz= >50 dB
20x: DC to 20 MHz (-3 dB)
20x: 17.5 ns
20x: @60Hz= >70 dB, @1 MHz= >40 dB
Maximum Input Voltage to Ground:
Maximum Differential Input Voltage:
(DC+AC peak) limit is determined by the point at which
the differential voltage probe starts clipping.
, 50 pF)
Note
Input Impedance:
Output Impedance: 50
Noise:
Offset: 10 mV into 1 M
Switch positions: OFF, 200x, 20x
Power
External:
Internal:
Power indicators:
Auto Stand By:
Environmental
Meets requirements of:
Temperature:
Altitude:
Meets requirements of:
Compliant with:
Max. Floating Output Voltage:
Between each input to shielding BNC: 5 M
Between the inputs: 10 M , 5 pF
200x: <2 mVrms
Via power adapter PM8907 (optional)
Battery Power: Alkaline 9V, IEC6LR61
Battery Life:
Green LED: ON at normal operation.
Red LED:
After 30 minutes, only when battery operated.
MIL-T-28800E, Type III, Class 3.
EN 50081-1, Electromagnetic Compatibility Generic Emission
Standard: EN55022 and EN60555-2.
EN 50082-2, Electromagnetic Compatibility Generic Immunity
Standard: IEC801-2, -3, -4, -5. (see also Tables 1 and 2)
This product is in conformity with Electromagnetic Compatibility
Directive 89/336/EEC and Low Voltage Directive 73/23/EEC.
This conformity is indicated by the symbol
Européenne”.
Operating: 0 C to +50 C (+32 F to +122 F)
Storage: -10 C to +60 C (+14 F to +140 F)
Operating: 3 km (9850 feet)
Storage: 12 km (40 000 feet)
EN61010-2-31 (IEC1010-2-31).
UL3111-1 (including listing)
CSA C22.2 No.1010.1-92 (including approval)
600V Category III, up to 400 Hz. (From shielding to ground.)
20x: <3 mVrms
Safety Specifications
Battery life measured @ 25 C with Duracell
battery. (Delivered with probe.)
To change from standby to normal operation turn switch
from OFF to 20x or 200x.
blinks at standby.
ON when battery needs to be replacing.
8 hour operation
400 hour in auto standby
, 6 pF
, i.e. “Conformité
®
alkaline
The DP120 conforms with the EEC directive 89/336 for EMC
immunity, as defined by IEC801-3, with the addition of the following
tables.
Servicing the DP120
For service information, contact your Fluke service center. To
locate an authorized service center, visit us on the World Wide
Web:
or call Fluke using any of the phone numbers listed below:
Available power adapters:
Service sheet:
Ordering Number: 4822 872 05374
Frequency range:
10 kHz to 25 MHz
20x
200x
Frequency range:
25 MHz to 1 GHz
20x
200x
Model
PM8907/801
PM8907/803
PM8907/804
PM8907/806
PM8907/807
PM8907/808
http://www.fluke.com
+1-888-993-5853 in U.S.A. and Canada
+31-402-678-200 in Europe
+1-425-356-5500 from other countries
The 230V rating of the PM8907/808 is not for use in
North America. A line plug adapter complying with the
applicable National Requirements may be provided to
alter the blade configurations for a specific country.
Description
Universal Europe 230V, 50 Hz
North America 120V, 60 Hz
United Kingdom 240V, 50 Hz
Japan 100V, 60 Hz
Australia 240V, 50 Hz
Universal 115V/230V
E = 3 V/m
<1%
<1%
E = 3 V/m
1%
1%
Table 1
Table 2
(in % of full dynamic range)
(in % of full dynamic range)
Note
Susceptibility
Susceptibility
E = 10 V/m
<1%
<1%
E = 10 V/m
2%
2%