D6F-20A5-000 Omron, D6F-20A5-000 Datasheet - Page 3

FLOW SENSOR MEMS 20L/MIN

D6F-20A5-000

Manufacturer Part Number
D6F-20A5-000
Description
FLOW SENSOR MEMS 20L/MIN
Manufacturer
Omron
Series
D6Fr
Type
MEMS Mass Flow Sensorr
Datasheets

Specifications of D6F-20A5-000

Sensing Range
0 ~ 20 LPM
Flow Sensor Type
Air
Voltage - Input
10.8 ~ 26.4 V
Port Size
0.196"(5mm)
Material - Body
Thermoplastic resin / Aluminum alloy
Accuracy
± 3%
Length/height, External
30mm
Flow Range
0 - 20 LPM
Output Type
Analog
External Width
78mm
Output Voltage Max
5.7V
Body Material
Thermoplastic Resin
External Depth
30mm
Flow Rate Min
0GPM (US)
Sensor Type
MEMS Air Flow
Rated Range
0 L/min to 20 L/min
Accuracy/reliability
+/- 3 % Full Scale
Analog Output
1 V to 5 V
Supply Voltage
12 V to 24 V
No. Of Ports
2
Sensor Output
Analog
Rohs Compliant
Yes
Air Volume
20l/min
Flow Rate Max
5.283GPM (US)
Accuracy %
3%
Lead Free Status / RoHS Status
Lead free / RoHS Compliant
Switch Function/rating
-
Lead Free Status / Rohs Status
Lead free / RoHS Compliant
Other names
D6F20A5000
Z2207
Test Results (typical performance)
D6F-20A@
(5 samples, repeated 10 times each)
Installation: D6F-@A5-000
Recommendation for right-angle, manifold mount connection;
Use M3 panhead screws for mounting, and tighten them to a maximum
torque of 0.59 N•m. Please seal INLET and OUTLET with ‘O’ rings.
Please install the product as shown below. Incorrect mounting may
cause product failure and make it impossible to obtain correct measure-
ments.
-0.2
-0.4
-0.6
-0.8
-1.0
1.0
0.8
0.6
0.4
0.2
0.0
0
2
4
6
5.8
8
Flow Rate (lpm)
10
12
1.9 ± 0.07
14
16
MEMS Airflow Sensor
18
Inlet
20
O ring
Recommended ‘O’ ring for sealing inlet and outlet orifice;
• Material: NBR70 (recommended for all applications including
• Example part number: JASO part number: CO 0003
• Alternative manufacturers with similar DO and W dimensions may
medical and food processing.)
be used
D6F-10A, -20A, -50A
Outlet
25