HD110C Amprobe, HD110C Datasheet - Page 2

Digital Multimeters STD HD DMM

HD110C

Manufacturer Part Number
HD110C
Description
Digital Multimeters STD HD DMM
Manufacturer
Amprobe
Type
DMMr
Datasheet

Specifications of HD110C

True Rms
No
Min / Max Hold
No
Peak Hold
No
Data Hold
Yes
Resistance Range
200 Ohms to 20 M Ohms
Equipment Type
Heavy Duty Digital Multimeter
Display Count
1999
Dmm Type
Hand Held
Voltage Measuring Range Dc
200mV To 1500V
Voltage Measuring Range Ac
200mV To 1000V
Current Measuring Range Dc
200µA To 2A
Current Measuring Range Ac
200µA To 2A
Battery Type
9V
Current, Range
200 μA, 2 mA, 20 mA, 200 mA, 2 A (Both AC and DC)
Dimensions
7.9 × 4.0 × 2.3 In.
Display Type
LCD
Material, Casing
Thermoplastic
Resistance, Range
200 Ω, 2 kΩ, 20 kΩ, 200 kΩ, 2 MΩ, 20 MΩ
Voltage, Range, Ac
200 mV, 2 V, 20 V, 200 V, 1000 V
Voltage, Range, Dc
200 mV, 2 V, 20 V, 200 V, 1500 V
Weight
20 Oz
Lead Free Status / RoHS Status
Lead free / RoHS Compliant
Lead Free Status / RoHS Status
na, Lead free / RoHS Compliant
Other names
2670779
HD110C Technical Specifications
General Specifications
Electrical Specifications
Amprobe® Test Tools
Display
Polarity Indication
Input overload indication
Low Battery Indication
Display Update Rate
Oper. Temp.
Storage Temp
Temperature coefficient
Altitude
Fuse
Power
Auto Power-Down
Battery Life
Dimensions, with holster
(H x W x D)
Weight (incl. battery)
Accessories
Case material
Safety
EMC
Function
DC Volts
Ranges
Resolution
Accuracy
Input Impedance
OL Protection
Transient Protection
AC Volts (45 Hz to 2 kHz)
Ranges
Resolution
Accuracy
Input Impedance
OL Protection
Transient Protection
DC Current
Ranges
Resolution
Accuracy
3-1/2 digit LCD, 1999 counts, with annunciators and menu features
Automatic
1
B; < 50 hours battery life remain, accuracy is no longer guaranteed
2.5/sec, nominal
0 °C to +50 °C, (< 75 % R.H.)
-20 °C to 60 °C, 0 to 80 % RH, battery removed
0.1 x (spec. accuracy)/°C (0 ° to 18 °C and 28° to 50 °C) all functions and ranges except for mVDC
0.3 x (spec. accuracy)/°C (0 ° to 18 °C and 28° to 50 °C) for mVDC
Up to 2000 m — indoor use
F 2 A/1500 V fuse (8 mm x 65 mm), I.R. 30 kA – Amprobe
Standard 9 V battery, NEDA 1604, JIS 006P,IEC 6F22
Meter powers down after approx. 30 min. of inactivity.
250 hours typical with carbon-zinc, 500 hours typical with alkaline
200 x 102 x 59 mm (7.9 in x 4.0 in x 2.3 in)
642 g (20 oz)
Heavy Duty CAT IV 1000 V Test leads with threaded alligator clips (TL1500), battery (in instrument), hex
wrench in holster, Magne-Grip flex strap and User Manual
Flame retardant, high-impact thermoplastic
Meets EN 61010-1 CAT IV 1000 V (1500 V DC max) Class II, EN 60529:IP67
Meets EN 61326-1
(Accuracies at 23 °C ± 5 °C, <75 % RH non-condensing, guaranteed for one year.)
Range
200 mV, 2 V, 20 V, 200 V, 1500 V
0.1 mV
All ranges
10 MΩ
1500 V dc/1000 V ac rms
12 kV impulse (1.2 µS/50 µS) based on EN 61010-1:2001 impulse requirement for at CAT IV 1000 V
product. This product should not be used in installations where transients exceed 12 kV.
200 mV, 2 V, 20 V, 200 V, 1000 V
0.1 mV
200 mV to 200 V (45-500 Hz)
200 mV to 200 V (500 Hz-2 kHz)
1000 V (50-60 Hz)
1000 V (60-500 Hz)
10 MΩ
1500 V dc/1000 V ac rms
12 kV impulse (1.2 µS/50 µS) based on EN 61010-1:2001 impulse requirement for at CAT IV 1000 V
product. This product should not be used in installations where transients exceed 12 kV.
200 µA, 2 mA, 20 mA, 200 mA, 2 A
0.1 µA
200 µA to 20 mA
200 mA to 2 A
CAT IV 1000 V (1500 V DC max) Class II, EN 60529:IP67
Accuracy
±(0.1 % rdg+2 dgts)
±(0.8 % rdg+4 dgts)
±(1.5 % rdg+5 dgts)
±(0.8 % rdg+4 dgts)
±(1.5 % rdg+5 dgts)
±(0.5 % rdg+2 dgts)
±(1.0 % rdg+2 dgts)
p/n FP700
Data Sheet