8808A-120 Fluke, 8808A-120 Datasheet - Page 3

Digital Multimeters 5.5 DIGIT MULTIMETER

8808A-120

Manufacturer Part Number
8808A-120
Description
Digital Multimeters 5.5 DIGIT MULTIMETER
Manufacturer
Fluke
Datasheet

Specifications of 8808A-120

Accuracy
0.01 %
Resistance Range
0 Ohms to 100 M Ohms
Lead Free Status / RoHS Status
Lead free / RoHS Compliant
Other names
2570235
Specifications are valid for 5-½ digit mode and after at least a half-hour warm-up.
DC Voltage Specifications
200 mV
2 V
20 V
200 V
1000 V
200 mV
2 V
20 V
200 V
1000 V
Range
Notes:
[1]
Notes:
[1]
Maximum Input ..........................................................1000 V on any range
Common Mode Rejection..........................................120 dB at 50 or 60 Hz 0.1
Normal Mode Rejection.............................................80 dB at Slow Rate
A/D Nonlinearity ........................................................15 ppm of range
Input Bias Current .....................................................<30 pA at 25 C
Settling Considerations.............................................Measurement settling times are affected by source impedance, cable
Range
Uncertainty given as
At some dual display measurements, the input impedance of 200 mV and 2 V ranges may be changed to 10 M .
199.999 mV
1.99999 V
19.9999 V
199.999 V
1000.00 V
(5-1/2 Digits)
Full-Scale
(% of reading + % of range)
0.01 + 0.003
0.01 + 0.002
0.01 + 0.003
0.01 + 0.002
0.01 + 0.002
23 °C
90 days
100 V
10 mV
10 V
1 mV
Slow
1 V
5 °C
Uncertainty
Input Characteristics
dielectric characteristics, and input signal changes
Electrical Specifications
[1]
Resolution
1000 V
Medium
100 mV
100 V
0.015 + 0.004
0.015 + 0.003
0.015 + 0.004
0.015 + 0.003
0.015 + 0.003
10 mV
10 V
23 °C
1 year
5°C
Fluke 8808A Digital Multimeter Extended Specifications Fluke Corporation 3
Fluke 8808A Digital Multimeter Extended Specifications
1000 V
(1 k
100 mV
100 V
10 mV
10 V
Fast
unbalance)
Temperature Coefficient/°C
Outside 18 – 28 °C
0.0015 + 0.0005
0.0020 + 0.0005
0.0015 + 0.0005
0.0015 + 0.0005
0.001 + 0.0005
Input Impedance
10 M
10 M
10 M
>10 G
>10 G
1 %
1 %
1 %
[1]
[1]
Fluke Corporation 3