FLUKE-116 Fluke, FLUKE-116 Datasheet - Page 2

Digital Multimeters HVAC TRUE RMS MULTIMETER

FLUKE-116

Manufacturer Part Number
FLUKE-116
Description
Digital Multimeters HVAC TRUE RMS MULTIMETER
Manufacturer
Fluke
Datasheet

Specifications of FLUKE-116

Accuracy
+/- 0.1 %
Equipment Type
Digital Multimeters
Display Count
6000
Ranging
Auto, Manual
Loz
helps prevent false readings due to ghost voltage
Lead Free Status / RoHS Status
Lead free / RoHS Compliant
Other names
2538803
Notes:
1
2
3
2 Fluke Corporation 116 HVAC Multimeter with Temperature and Microamps
Accuracy specifications
All ac voltage ranges except Auto-V/LoZ are specified from 1 % to 100 % of range. Auto-V/LoZ is specified from 0.0 V.
Because inputs below 1 % of range are not specified, it is normal for this and other true-rms meters to display non-
zero readings when the test leads are disconnected from a circuit or are shorted together. For volts, crest factor of ≤ 3
at 4000 counts, decreasing linearly to 1.5 at full scale. AC volts is ac-coupled. Auto-V LoZ, and ac mV are dc-coupled.
Temperature uncertainty (accuracy) does not include the error of the thermocouple probe.
Frequency is ac coupled, 5 Hz to 50 kHz for ac voltage. Frequency is dc coupled, 45 Hz to 5 kHz for ac current.
Measurement
DC millivolts
DC volts
Auto volts
AC millivolts
true-rms
AC volts
Continuity
Ohms
Diode test
Capacitance
Lo-Z capacitance
Temperature
(Type-K
thermocouple)
AC µamps
true-rms
(45 Hz to 500 Hz)
DC µamps
Hz (V or A input)
1
true-rms
1
2
3
Range
600.0 mV
6.000 V
60.00 V
600.0 V
600.0 V
600.0 mV
6.000 V
60.00 V
600.0 V
600 Ω
600.0 Ω
6.000 kΩ
60.00 kΩ
600.0 kΩ
6.000 MΩ
40.00 MΩ
2.000 V
1000 nF
10.00 µF
100.0 µF
9999 µF
100 µF to 1000 µF
> 1000 µF
1 nF to 500 µF
-40 °C to 400 °C
-40 °F to 752 °F
600.0 µA
600.0 µA
99.99 Hz
999.9 Hz
9.999 kHz
50.00 kHz
Resolution
0.1 mV
0.001 V
0.01 V
0.1 V
0.1 V
0.1 mV
0.001 V
0.01 V
0.1 V
1 Ω
0.1Ω
0.001 kΩ
0.01 kΩ
0.1 kΩ
0.001 MΩ
0.01 MΩ
0.001 V
1 nF
0.01 µF
0.1 µF
1 µF
0.1 °C
0.2 °F
0.1 µA
0.1 µA
0.01 Hz
0.1 Hz
0.001 kHz
0.01 kHz
1.9 % + 2
5 % + 20
Accuracy ± ([% of reading] + [counts])
0.5 % + 2
0.5 % + 2
2.0 % + 3 (dc, 45 Hz to 500 Hz)
4.0 % + 3 (500 Hz to 1 kHz)
1.0 % + 3 (dc, 45 Hz to 500 Hz)
2.0 % + 3 (500 Hz to 1 kHz)
1.0 % + 3 (45 Hz to 500 Hz)
2.0 % + 3 (500 Hz to 1 kHz)
Beeper on < 20 Ω off > 250 Ω; detects
opens or shorts of 500 µs or longer.
0.9 % + 2
0.9 % + 1
1.5 % + 2
0.9 % + 2
1.9 % + 2
10 % + 2 typical
1 % + 10
1 % + 18
1.5 % + 3 (2.5 % + 3 > 500 Hz)
1.0 % + 2
0.1 % + 2
2
2
Fluke. Keeping your world
Fluke Corporation
PO Box 9090, Everett, WA USA 98206
Fluke Europe B.V.
PO Box 1186, 5602 BD
Eindhoven, The Netherlands
For more information call:
In the U.S.A. (800) 443-5853 or
Fax (425) 446-5116
In Europe/M-East/Africa (31 40) 2 675 200 or
Fax (31 40) 2 675 222
In Canada (800)-36-FLUKE or
Fax (905) 890-6866
From other countries +1 (425) 446-5500 or
Fax +1 (425) 446-5116
Web access: http://www.fluke.com
©2006 Fluke Corporation. All rights reserved.
Specifications subject to change without notice.
Printed in U.S.A. 8/2006 2634037 D-EN-N Rev B
Ordering
information
Fluke-116
Included
TL75 Test Leads, 80BK Integrated
Temperature Probe, holster, User’s
manual and 9 V battery (installed).
up and running.
HVAC Multimeter
with Temperature
and Microamps