33XR-A Amprobe, 33XR-A Datasheet - Page 3

Digital Multimeters PROF DMM W/TEMP AND CAPACITANCE

33XR-A

Manufacturer Part Number
33XR-A
Description
Digital Multimeters PROF DMM W/TEMP AND CAPACITANCE
Manufacturer
Amprobe
Datasheet

Specifications of 33XR-A

Accuracy
+/- 0.7 %
True Rms
No
Min / Max Hold
Yes
Peak Hold
Yes
Data Hold
Yes
Resistance Range
400 Ohms to 4 M Ohms
Capacitance Range
4 uF to 4000 uF
Frequency
40 MHz
Equipment Type
Digital Multimeter with Temperature and Capacitance
Display Count
3999
Ranging
Manual
Dmm Type
Hand Held
Voltage Measuring Range Dc
400mV To 1000V
Voltage Measuring Range Ac
400mV To 750V
Current Measuring Range Dc
400µA To 10A
Current Measuring Range Ac
400µA To 10A
Lead Free Status / RoHS Status
Lead free / RoHS Compliant
Lead Free Status / RoHS Status
na, Lead free / RoHS Compliant
Other names
2727788
Specifications
Capacitance
Temperature
Frequency (autoranging)
Continuity
Diode Test
Peak Hold
Overload Protection
33XR-A Digital Multimeter
Function
Ranges
Accuracy
Resolution
Test Voltage
Test Frequency
Input Protection μA/mA input:
Ranges
Accuracy
Resolution
Ranges
Resolution
Sensitivity
Min Pulse Width
Duty Cycle Limits
Audible Indication
Response Time
Test Current
Resolution
Open Circuit Volts
AC Volts Accuracy
AC Current Accuracy
Response Time
AC/DC Voltage
AC/DC Current
Resistance
Continuity
Diode Test
Capacitance
Temperature
Frequency
(23 °C ± 5 °C, <75 % R.H. non-condensing), cont.
Range
4 μF, 40 μF, 400 μF, 4000 μF
4 μF range
40 μF, 400 μF ranges
4000 μF range
0.1 μF
< 3.0 V
10 Hz
0.315 A/1000 V fast blow ceramic fuse 6.3 x 32 mm
-20 °C to 1000 °C, -4 °F to 1832 °F
-20 °C to 10 °C
10 °C to 200 °C
200 °C to 1000 °C
-4 °F to 50 °F
50 °F to 400 °F
400 °F to 1832 °F
1 °C, 1 °F
4 kHz, 40 kHz, 400 kHz, 4 MHz, 40 MHz
1 Hz
10 Hz to 4 MHz: > 1.5 V AC RMS
4 MHz to 40 MHz: > 2 V AC RMS, < 5 V AC RMS
> 25 ns
> 30 % and < 70 %
< 35 W
100 ms
1.0 mA (approximate)
1 mV
3.0 V DC typical
40 V to 750 V
> 1 ms
400 mV range: 1000 V DC or 750 V AC RMS 15 sec
All other ranges: 1000 V DC or 750 V AC RMS
uA/mA input: 0.315 A /1000 V fast-blow ceramic (6.3 x 32 mm)
10 A input: 10 A /1000 V fast-blow ceramic (10 x 38 mm)
(10 A for 4 minutes maximum followed by a 12 minute cooling period)
1000 V DC OR 750 V AC RMS
1000 V DC OR 750 V AC RMS
1000 V DC OR 750 V AC RMS
0.315 A /1000 V fast-blow ceramic (6.3 x 32 mm)
1000 V DC OR 750 V AC RMS
1000 V DC OR 750 V AC RMS
www.Amprobe.com
Accuracy
± (5.0 % rdg + 10 dgts)
± (5.0 % rdg + 5 dgts)
± (5.0 % rdg + 15 dgts)
± (2.0 % rdg + 4 °C )
± (1.0 % rdg + 3 °C )
± (3.0 % rdg + 2 °C )
± (2.0 % rdg + 8 °F)
± (1.0 % rdg + 6 °F)
± (3.0 % rdg + 4 °F)
± (0.1 % rdg + 3 dgts)
± (1.5 % rdg + 3 dgts)
± (3.0 % rdg + 60 dgts)
± (3.0 % rdg + 60 dgts)
Data Sheet