HD160C Amprobe, HD160C Datasheet - Page 3

Digital Multimeters HD TRUE RMS

HD160C

Manufacturer Part Number
HD160C
Description
Digital Multimeters HD TRUE RMS
Manufacturer
Amprobe
Datasheet

Specifications of HD160C

Accuracy
+/- 2 %
True Rms
Yes
Min / Max Hold
No
Peak Hold
No
Data Hold
No
Resistance Range
40 Ohms
Capacitance Range
40 nF to 400 uF
Frequency
0.7 Hz to 1.3 Hz
Equipment Type
Heavy Duty Digital Multimeter
Display Count
10000
Ranging
Auto
Dmm Type
Hand Held
Voltage Measuring Range Dc
1000mV To 1500V
Voltage Measuring Range Ac
1000mV To 1000V
Current Measuring Range Dc
100µA To 2A
Current Measuring Range Ac
100µA To 2A
Lead Free Status / RoHS Status
Lead free / RoHS Compliant
Lead Free Status / RoHS Status
na, Lead free / RoHS Compliant
Other names
2670787
HD160C Technical Specifications (continued)
DC Current
AC Current True rms
Resistance
Continuity Test
Capacitance
Temperature
Function
Ranges
Resolution
Accuracy
Input protection
Burden voltage
OL Protection
Ranges
Resolution
Accuracy (45 Hz to 1kHz)
Voltage burden
Conversion type
Crest factor
OL protection
Ranges
Resolution
Accuracy
Overload protection, all ranges 1500 V dc or 1000 V ac rms
Audible indication
Response time
Overload protection
Ranges
Resolution
Accuracy
Test voltage
Test Frequency
Overload protection
Ranges
Resolution
Accuracy
Overload protection
Current True rms
Frequency
voltage
burden
protection
Range
100 µA, 1000 µA, 10mA, 100mA, 400mA, 2A (Auto/Manual ranging)
0.01µA in 100 µA range
100 µA range
1000 µA to 400 mA ranges
2 A range
2 A/1500 V fast blow ceramic fuse 8×65 mm on A input , FP700
µA range of 1 mV/1 µA, mA range of 10 mV/1 mA, 2A range of 500 mV/1A, 500 mV max. (2 V max. on
1000 µA, 100 mA, 400 mA, 2 A ranges)
A Input (F 2 A/1500 V, size 8 x 65 mm IR fast blow ceramic)
100 µA, 1000 µA, 10 mA, 100 mA, 400 mA, 2 A
0.01 µA in 100 µA range
100 µA to 100 mA
400 mA
2 A
see DC Current
True rms ac coupled 10 to 100% of range
≤ 3
see DC Current
1000 Ω, 10 kΩ, 100 kΩ, 1000 kΩ, 10 MΩ, 40 MΩ
0.1 Ω in 1000 Ω range
1000 Ω to 1000 kΩ ranges
10 MΩ range
40 MΩ range
Less than 40 Ω
100 ms
1500 V dc or 1000 V ac rms
40 ηF, 400 ηF, 4 µF, 40 µF, 400 µF (3999 counts) (Auto/Manual ranging)
0.01 ηF
40 ηF, 400 µF ranges
400 ηF to 40 µF ranges
< 1 V
1.3 Hz on 40 ηF to 40 µF ranges; 0.7 Hz on 400 µF range
1500 V dc or 1000 V ac rms
-20 °C ~ 1300 °C (-4° F ~ 2372 °F)
1 °C, 1 °F
-20 °C ~ 10 °C
10 °C ~ 200 °C
200 °C ~ 1300 °C
-4 °F ~ 50 °F
50° F ~ 400 °F
400 °F ~2372 °F
1500 V dc or 1000 V ac rms
DC Current
dc or 1000 V ac rms
or 1000 V ac rms
ac rms
Accuracy
± (0.5% rdg + 10 dgts)
± (0.5% rdg + 5 dgts)
± (1.5% rdg + 10 dgts)
± (1.5 % + 10 dgts)
± (2.0 % + 10 dgts)
± (2.5 % + 20 dgts)
± (0.5 % rdg + 8 dgts)
± (1.0 % rdg + 10 dgts)
± (2.0 % rdg + 10 dgts)
±(3.0% rdg +10 dgts)
±(3.0% rdg +5 dgts)
3999 counts
±(2.0% rdg + 4 °C)
±(1.0% rdg + 3 °C)
±(2.0% rdg + 2 °C)
±(2.0% rdg + 8 °F)
±(1.0% rdg + 6 °F)
±(2.0% rdg + 4 °F)
www.Amprobe.com
(continued on next page)
Data Sheet