DMM4050 Tektronix, DMM4050 Datasheet

Digital Multimeters 6.5 DIGIT (24 PPM) PRECISION METER USB

DMM4050

Manufacturer Part Number
DMM4050
Description
Digital Multimeters 6.5 DIGIT (24 PPM) PRECISION METER USB
Manufacturer
Tektronix
Datasheet

Specifications of DMM4050

Accuracy
0.0024 %
True Rms
Yes
Resistance Range
10 Ohms to 1 GOhms
Frequency
47 Hz to 440 Hz
Equipment Type
Digital Multimeter
Display Count
6.5
Current, Rating
100 μA to 10 A, 100 pA Resolution
Dimensions
3.46 × 8.56 × 11.7 (88 × 217 × 297) In. (mm)
Display Type
Vacuum Fluorescent Display, dot matrix
Frequency Range
47 to 440 Hz
Includes
National Instrument's LabVIEW SignalExpress™ TE Limited Ed.
Power Consumption
28 VA peak (12 W average)
Resolution
6.5 Digit Resolution
Temperature, Operating
0 to 55 °C
Temperature, Storage
–40 to 70 °C
Voltage, Rating
100 mV to 1000 V, 100 nV Resolution
Warranty
3 Years
Weight
8.0 (3.6) Lb. (kg)
Lead Free Status / RoHS Status
Lead free / RoHS Compliant
Digital Multimeters
Tektronix DMM4050 and DMM4040 Data Sheet
DMM4050
Features & Benefits
Key Performance Specifications
6.5 Digit Resolution
Basic VDC Accuracy of up to 0.0024% (1 yr.)
100 mV to 1000 V Voltage Range, with up to 100 nV Resolution
100 μA to 10 A Current Range, with up to 100 pA Resolution
10 Ω to 1 GΩ Ohm Range, with up to 10 μΩ Resolution
CAT I 1000 V, CAT II 600 V
Available Functions and Features
Connectivity
3-year Warranty
Volts, Ohms, and Amps Measurements
Diode and Continuity Testing
Frequency and Period Measurements
Temperature and Capacitance Measurements (DMM4050)
2×4 Ohms 4-wire Measurement Technique
Trendplot™ Paperless Data Recorder Mode
Measurement Statistics
Histogram Mode
Front and Rear 2×4 Measurement Inputs
USB Host Port on Front Panel for Easy Storage of Measurement Data
and Instrument Settings
RS-232, LAN, and GPIB Ports on Rear Panel for Quick PC Connectivity
Includes USB to RS-232 Interface Adapter Cable
Includes National Instrument’s LabVIEW SignalExpress™ TE Limited
Edition for Connecting Your Bench

DMM4050 Summary of contents

Page 1

... Digital Multimeters Tektronix DMM4050 and DMM4040 Data Sheet DMM4050 Features & Benefits Key Performance Specifications 6.5 Digit Resolution Basic VDC Accuracy 0.0024% (1 yr.) 100 mV to 1000 V Voltage Range, with up to 100 nV Resolution 100 μ Current Range, with up to 100 pA Resolution 10 Ω ...

Page 2

... A USB port on the front panel makes it easy to store measurement data and instrument setups to a memory stick. Or connect to your PC with the LAN, RS-232, or GPIB port on the back panel. A USB to RS-232 interface adapter cable is included standard with the DMM4050/4040 to make it easy to connect to your PC’s USB port. Simple and Accurate 4-wire Measurements Patented split terminal jacks for the 2× ...

Page 3

... PC. Only Tektronix offers a connected test bench of intelligent instruments to simplify and speed debug of your complex design. Performance You Can Count On In addition to industry-leading service and support, every DMM4050 and DMM4040 multimeter comes backed with a three-year standard warranty ...

Page 4

... Triggering Characteristic Samples per Trigger Trigger Delay External Trigger Delay External Trigger Jitter Trigger Input Trigger Output Memory Model DMM4050/4040 in. 3.46 Math Functions 8.56 Zero, dBm, dB, MX+B, Offset, DCV ratio and TrendPlot, Histogram, Statistics 11.7 (min/max/average/standard deviation), and Limit Test lb. Electrical 8 ...

Page 5

... V 1.000000 10.00000 V 100 V 100.0000 V 1000 V 1,000.000 V * Inputs beyond ±14 V are clamped through 200 kΩ typical. 10 MΩ is default input impedance. 2 Digital Multimeters — Tektronix DMM4050 and DMM4040 Resolution 4½ Digits 5½ Digits 10 μV 1 μV 100 μV 10 μ 100 μ 100 mV ...

Page 6

... Data Sheet DMM4050 Accuracy Accuracy is given as ±(% measurement + % of range) Range 24 Hour (23 °C ±1 °C) 100 mV 0.0025 + 0.003 1 V 0.0018 + 0.0006 10 V 0.0013 + 0.0004 100 V 0.0018 + 0.0006 1000 V 0.0018 + 0.0006 DMM4040 Accuracy Accuracy is given as ±(% measurement + % of range) Range 24 Hour (23 °C ±1 °C) 100 mV 0 ...

Page 7

... Crest factor 4-5, 0.5% of full scale Only applies for non-sinusoid signals Input Characteristics Range Resolution 100 mV 100.0000 1.000000 10.00000 V 100 V 100.0000 V 1000 V 1,000.000 V Digital Multimeters — Tektronix DMM4050 and DMM4040 Volts-Hertz product 7 Resolution 4½ Digits 5½ Digits 10 μV 1 μV 100 μV 10 μ 100 μ 100 mV ...

Page 8

... Data Sheet DMM4050/4040 Accuracy Accuracy is given as ±(% measurement + % of range) Range Frequency 100 mV 3 – – – 20 kHz 20 – 50 kHz 50 – 100 kHz 100 – 300 kHz – – – 20 kHz 20 – 50 kHz 50 – 100 kHz 100 – 300 kHz – – – ...

Page 9

... GΩ 1.0 + 0.01 Additional Ohms Errors Digits 6½ 6½ 5½ 5½ 4½ Digital Multimeters — Tektronix DMM4050 and DMM4040 Resolution 4½ Digits 5½ Digits 1 mΩ 100 μΩ 10 mΩ 1 mΩ 100 mΩ 10 mΩ 1 Ω ...

Page 10

... A * Part range Available on the front-panel terminal only 400 mA available in software version 2.0 or greater only. 400 mA continuously; 550 mA for 2 minutes on, 1 minute off. 7 DMM4050/4040 Accuracy Accuracy is given as ±(% measurement + % of range) Range 24 Hour (23 °C ±1 °C) 100 μA 0. 0.007 + 0.005 10 mA ...

Page 11

... Part range Available at front-panel connectors only 400 mA available in software version 1.0.700.18 or greater only. 400 mA continuously; 550 mA for 2 minutes on, 1 minute off; maximum crest factor 3:1 at 400 mA. 9 Digital Multimeters — Tektronix DMM4050 and DMM4040 Resolution 4½ Digits 5½ Digits 100 μA 100 nA 10 μ ...

Page 12

... Data Sheet DMM4050/4040 Accuracy Accuracy is given as ±(% measurement + % of range) Range Frequency 100 μA 3 – – – 5 kHz 5 – 10 kHz – – – 5 kHz 5 – 10 kHz – – – 5 kHz 5 – 10 kHz 100 mA 3 – – – 5 kHz 5 – 10 kHz 400 mA* 3 – ...

Page 13

... Input >100 mV. For 10 – 100 mV, multiply percent measurement error by 10. Gate Time vs. Resolution Resolution Gate Time 0.01 5½ 0.1 6½ 1.0 6½ Digital Multimeters — Tektronix DMM4050 and DMM4040 24 Hour 90 Days (23 °C ±1 °C) (23 °C ±5 °C) 0.1 0.1 0.05 0.05 0.03 ...

Page 14

... Data Sheet Capacitance (DMM4050 Only) Accuracy is stated as ±(% of measurement + % of range) Range 100 nF 1 μF 10 μF 100 μ 100 mF * Stated accuracy is attained when Zero function is used. 12 Temperature (DMM4050 only) Test Current Accuracy is stated as ± °C and is based on a Platinum RT100 (DIN IEC 751, 385 type) RTD with less than 10 Ω lead resistance. The accuracy listed in the table below are valid only when using the 4-wire RTD measurement function. Specifi ...

Page 15

... Speeds available in OutG SW 1.0.700.18 or higher. Note that the measurement rates for RS232 can vary depending on the baud rate chosen. If the baud rate selected is 115,200, the maximum measurement rate is 711 measurements. The LAN bus has a maximum measurement rate of 963 measurements. Digital Multimeters — Tektronix DMM4050 and DMM4040 Diode Test ...

Page 16

... Digit Multimeter DMM4050 6.5 Digit Multimeter DMM4050/4040 Includes: Meter, TL710 test leads, line cord, spare line fuse, statement of cal practices, Warranty statement, Safety and Installation Guide, CD-ROM with user manual (English, French, Italian, German, Spanish, Simplified Chinese, Traditional Chinese, Korean, Russian, Japanese), RS-232 to USB Adapter Cable, National Instruments LabVIEW SignalExpress™ ...

Page 17

... Digital Multimeters — Tektronix DMM4050 and DMM4040 www.tektronix.com/dmm4050_4040 17 ...

Page 18

... Data Sheet 18 www.tektronix.com/dmm4050_4040 ...