DMM4040 Tektronix, DMM4040 Datasheet - Page 13
DMM4040
Manufacturer Part Number
DMM4040
Description
Digital Multimeters 6.5 DIGIT (35 PPM) PRECISION METER USB
Manufacturer
Tektronix
Datasheet
1.DMM4040.pdf
(20 pages)
Specifications of DMM4040
Accuracy
0.0035 %
True Rms
Yes
Resistance Range
10 Ohms to 1 GOhms
Frequency
47 Hz to 440 Hz
Equipment Type
Digital Multimeter
Display Count
6.5
Current, Rating
100 μA to 10 A, 100 pA Resolution
Dimensions
3.46 × 8.56 × 11.7 (88 × 217 × 297) In. (mm)
Display Type
Vacuum Fluorescent Display, dot matrix
Frequency Range
47 to 440 Hz
Includes
National Instrument's LabVIEW SignalExpress™ TE Limited Ed.
Power Consumption
28 VA peak (12 W average)
Resolution
6.5 Digit Resolution
Temperature, Operating
0 to 55 °C
Temperature, Storage
–40 to 70 °C
Voltage, Rating
100 mV to 1000 V, 100 nV Resolution
Warranty
3 Years
Weight
8.0 (3.6) Lb. (kg)
Lead Free Status / RoHS Status
Lead free / RoHS Compliant
Frequency
Characteristic
Gate Time
Measurement
Method
Settling
Considerations
Measurement
Considerations
DMM4050/4040 Accuracy
*
*
Gate Time vs. Resolution
Accuracy is given as ± % measurement
Range
100 mV to 1000 V*
Gate Time
0.01
0.1
1.0
10
11
Limited to 8 × 10
Input >100 mV. For 10 – 100 mV, multiply percent measurement error by 10.
7
Volt-Hertz
10, 11
Description
Programmable to 1 s, 100 ms, and 10 ms
Flexible counting technique. AC-coupled input using
the AC voltage measurement function
When measuring frequency or period after a DC
offset voltage change, errors may occur. For the
most accurate measurement, wait up to 1 second for
the input blocking capacitor to settle
To minimize measurement errors, shield inputs
from external noise when measuring low-voltage,
low-frequency signals
Resolution
5½
6½
6½
300 kHz – 1 MHz
40 Hz – 300 kHz
Frequency
10 – 40 Hz
5 – 10 Hz
3 – 5 Hz
(23 °C ±1 °C)
24 Hour
0.006
0.006
0.05
0.03
0.1
Additional Low-frequency Errors
Error stated as percent of measurement for inputs >100 mV. For 10 – 100 mV,
multiply percent by 10.
Frequency
3 – 5 Hz
5 – 10 Hz
10 – 40 Hz
40 – 100 Hz
100 – 300 Hz
300 Hz – 1 kHz
>1 kHz
(23 °C ±5 °C)
Digital Multimeters — Tektronix DMM4050 and DMM4040
90 Days
0.05
0.03
0.01
0.01
0.1
6½
0
0
0
0
0
0
0
(23 °C ±5 °C)
1 Year
0.05
0.03
0.01
0.01
0.1
www.tektronix.com/dmm4050_4040
NPLC
0.12
0.17
0.06
0.03
0.01
5½
0.2
0
Outside 18 to 28 °C
Coefficient/°C
Temperature
0.005
0.005
0.001
0.001
0.001
0.12
0.17
0.21
0.21
0.07
0.02
4½
0.2
13