I1010 Fluke, I1010 Datasheet

Clamp Multimeters & Accessories AC/DC CURRENT CLAM

I1010

Manufacturer Part Number
I1010
Description
Clamp Multimeters & Accessories AC/DC CURRENT CLAM
Manufacturer
Fluke
Type
Current Clampr
Datasheet

Specifications of I1010

Accuracy
+/- 2 %
Frequency
400 Hz
Voltage
1000 A
Ranging
1000 A
Features
60 hour battery life
Bandwidth
10 kHz
Conductor Size
1.18 in.
Current, Rating
0.5 to 1000 A
Dimensions
3.09 in. L x 8.21 in. W x 1.87 in. H
Temperature Coefficient
0 to 18 °C
Temperature, Operating
-10 to +50 °C
Voltage, Rating
600 V (RMS)
Lead Free Status / RoHS Status
Lead free / RoHS Compliant
Other names
617735
P
PN 617677 September 1996, Rev.1, 4/97
1996, 1997 Fluke Corporation. All rights reserved. Printed in Thailand.
Avoid working alone to ensure that assistance can be rendered.
Do not use the clamp if the clamp or leads appear damaged, or if you
suspect that the clamp is not operating properly.
Use extreme caution when working around bare conductors or bus
bars. Accidental contact with the conductor could result in electric
shock.
Use caution when working with voltages above 30V rms or 60V dc.
Such voltages pose a shock hazard.
Never use the clamp on a circuit with a voltage higher than 600V
(CAT. III) or a frequency higher than 400 Hz (i410) or 2 kHz (i1010.)
The clamp may be damaged if these limits are exceeded.
Use the clamp only as specified in this instruction sheet; otherwise,
the protection provided by the clamp may be impaired.
Read the operating instructions before use and follow all safety
information.
CAT. III - Industrial Locations (IEC 1010-1).
AC/DC Current Clamp
Read First: Safety Information
i410/i1010
Instruction Sheet
®

I1010 Summary of contents

Page 1

... Such voltages pose a shock hazard. Never use the clamp on a circuit with a voltage higher than 600V (CAT. III frequency higher than 400 Hz (i410 kHz (i1010.) The clamp may be damaged if these limits are exceeded. Use the clamp only as specified in this instruction sheet; otherwise, the protection provided by the clamp may be impaired ...

Page 2

Features and Connections CAT 600V 600A A ZERO i410 CLAMP METER 600V Minimum voltmeter requirements: 76 TRUE RMS MULTIMETER V ar001i.eps Jaw Lever Jaw Centering Marks ON Indicator ON/OFF Switch Zero Adjust Battery Access Output Cable Accepts safety- shrouded banana ...

Page 3

... Measuring DC Current Maximum: 400A dc (i410) or 1000A dc (i1010) Connect to voltmeter. Select mV dc. Set ON. Adjust ZERO (jaws empty.) Clamp and center around conductor. Read voltmeter ( 1A.) CAT CAT 600V 600A A ZERO i410 CLAMP METER 600V 600V 600A A ZERO i410 CLAMP METER 600V 76 TRUE RMS MULTIMETER mV ar002i ...

Page 4

... Periodically wipe the case with a damp cloth and mild detergent. Do not use abrasives or solvents. Service and Parts The Current Clamp should be serviced only by a qualified service technician. For service information, contact your nearest Fluke dealer or service center. Specifications Operating Temperature Storage Temperature Relative Humidity ...

Page 5

... Hz, Crest Factor 400A) 3 kHz i1010 1 - 600A ac rms * 1 - 1000A dc 0.5 - 1000A 2.0% + 0.5A (0- 1000A) 2.0% + 0.5A 400 Hz, Crest Factor 3. 3.0% + 0.5A, 400 kHz sine wave 600A) ...