40027 Simpson, 40027 Datasheet

no-image

40027

Manufacturer Part Number
40027
Description
Analog Multimeters 228 IEC AC LEAKAGE
Manufacturer
Simpson
Datasheet

Specifications of 40027

Current Rating
228 A
Frequency
60 Hz
Measured Components
AC/DC Milliameter
Resistance
1 MOhms
Display Type
228 Leakage Current Tester
Accuracy
± 2% of Reading @ 60 Hz (Output)
Measuring Range
0.3 to 10 MIU (Reaction Response Network)
Size
7 in. x 5.25 in. x 3.125 in.
Standards
IEC Approved, ANSI and UL Listed
Lead Free Status / RoHS Status
Lead free / RoHS Compliant
Designed specifically to measure hazardous "Leakage Currents" or "Touch Currents"
which may appear when a person touches the conductive surface of electrical or
electronic equipment. The 228 is designed around IEC990, ANSI and UL guidelines.
Model 228 Special Purpose AC/DC Milliameter
• Reads in Measurement Indication Units (MIU) up to .005 MIU
• Output Allows Measurement of Peak Current in Non-Sinusoidal Wave Forms
• Let-Go, Reaction, and Burn Hazard Response Networks
• Detects dangerous AC & DC Leakage currents
• Includes Test Leads, Alligator Clips, Batteries and Manual
The specifications apply to sinusoidal AC waveforms only. Accuracy is not guaranteed for
non-sinusoidal or complex waveforms.
Specifications
Any discussion in this document regarding UL, ANSI or IEC specifications is for Reference purposes only. The input network utilized in the M228 is detailed in
Figures 1 through 3 on page 4 of the Manual. The customer is advised to obtain the latest specification from the rating agency. NOTE: For specification
information call ANSI at (212) 642-4900 or UL in Northbrook, IL at (847) 272-8800.
• True RMS Readings
Equivalent network component values:
Equivalent measuring instrument load:
Ranges:
Current accuracy:
Meter measurement method:
Meter frequency response:
Output sensitivity:
Output accuracy:
Output frequency response:
Voltmeter range:
Voltmeter accuracy:
Voltmeter frequency response:
Power requirements:
Physical:
Environmental:
Maximum operating range:
Ordering Information
Leakage Tester
228 Leakage Current Tester
GENERAL
D
2Hz to 19Hz (Accuracy not supported)
20 Hz to 200 KHz
200 KHz to 1 MHz
DC to 50Hz
50 Hz to 100 KHz
100 kHz to 1 MHz
t C
1 o
z H
2Hz to 19Hz (Accuracy not supported)
20 Hz to 1KHz
D
t C
1 o
z H
Catalog No.
40027
ACCESSORIES
Test Leads w/Screw-On Alligator Clips
Case, Black Padded Nylon Carrying
Specifications subject to change without notice.
Response Network
1 M - 10pF
0.3, 1, 3, 10 M.I.U.
Full scale meter indication equals 1V RMS (Measured with a 1 M , 12 pF load)
2% F.S.
2% F.S.
2.5% of reading
Reaction
Designed around IEC 990 (excluding fuse)
(Relative to ANSI C101-1992 or UL-1459 2nd edition)
(Relative to ANSI C101-1992 or UL-1459 2nd edition)
7" x 5.25" x 3.125", 2-1/2 lbs (1.4kg), ABS plastic, fused input
228 Leakage Current Tester
27°C
P
P
n i o
n i o
2% of reading @ 60 Hz
r e t
r e t
Catalog No.
h s
h s
2° C, 70% non condensing relative humidity.
l l a
l l a
Response Network
1 M - 62 pF
0.3, 1, 3, 10 M.I.U.
+2% /-3% of reading
0-300 V (AC or DC)
DC to 1 KHz
0° to 40° C
00125
00834
a r t
a r t
2% F.S. @ 60 Hz
2.5% of F.S.
2.5% F.S.
2% of reading
5% of reading
3% F.S. @ 60 Hz (Add
(2) 9V (NEDA 1604A) batteries
k c
k c
True RMS
Let-Go
t i w
t i w
n i h
n i h
Optional carrying case
includes convenient
storage pouch for test
leads and Operators
Manual.
% 5
% 5
f o
f o
e p
e p
k a
k a
1% for every additional 100Hz)
Unweighted Burn Hazard
Response Network
1 M
0-100 mA RMS
2% of reading
- 1 pF
2% F.S.
5% F.S.
rev_3 12/10