FLUKE-1503 Fluke, FLUKE-1503 Datasheet - Page 3

Insulation Testers / Megohmmeters INSULATION TESTER

FLUKE-1503

Manufacturer Part Number
FLUKE-1503
Description
Insulation Testers / Megohmmeters INSULATION TESTER
Manufacturer
Fluke
Datasheet

Specifications of FLUKE-1503

Resistance
0.01 to 20 KOhms
Display Type
LCD Display
Current Rating
1 mA
Measured Components
Insulation Resistance
Safety Rating
600 V, CAT IV
Included Accessories
Remote Probe, Test Leads and Probes, Alligator Clips
Lead Free Status / RoHS Status
Lead free / RoHS Compliant
Other names
2427883
Digital Multimeters
Fluke 1507/1503 Insulation Resistance Testers
AC/DC Voltage Measurement
Accuracy
Input impedance
Common mode rejection
ratio (1 kΩ unbalanced):
Overload Protection
Earth bond resistance measurement
Range/Resolution
Accuracy
Overload Protection
Open Circuit Test Voltage
Short circuit current
Insulation specifications
Measurement Range
Test voltage
Test Voltage Accuracy
Short circuit current
Auto Discharge
Live circuit indicator
Maximum Capacitive
Load
Measure Accuracy:
General Specifications
Maximum voltage
applied to any
terminal:
Storage
Temperature
Operating
Temperature
Temperature
>
600 V ac rms or dc
-40 °C to 60 °C (-40 °F to 140 °F)
-20 °C to 55 °C (-4 °F to 131 °F)
0.05 x (specified accuracy) per °C for temperatures < 18 °C or > 28 °C (< 64 °F
Range
Resolution
50 Hz to 400 Hz ± (% of Rdg + Digits) ± (2 % + 3)
1507: 0.01 MΩ to 10 GΩ
1503: 0.01 MΩ to 2000 MΩ
1507: 50 V, 100 V, 250 V, 500 V, 1000 V
1503: 500 V, 1000V
50 V:
100 V: ±(3 % + 5)
250 V: ±(1.5 % + 5)
500 V: ±(1.5 % + 5)
1000 V: ±(1.5 % + 5) to 2000 MΩ, ±(10 % + 3) above 2000 MΩ
20.00 Ω 0.01 Ω
200.0 Ω 0.1 Ω
2000 Ω
20.00 kΩ 0.01 kΩ
3 MΩ (nominal), < 100 pF
> 60 dB at dc, 50 or 60 Hz
600 V rms or dc
+ 20 %, - 0 %
1 mA nominal
Discharge time < 0.5 second for C = 1 µF or less
Inhibit test if terminal voltage > 30 V prior to initialization of test
Operable with up to 1 µF load
± (1.5 % + 3)
2 V rms or dc
> 4.0 V, < 8 V
> 200.0 mA
±(3 % + 5)
1.0 Ω
600.0 V
0.1 V