The 724 Workstation Monitor cannot be re-calibrated after the initial factory calibration.
3M recommends that the operation of the monitor be verified at a 1 year interval or more
frequently if used in harsh environments such as high humidity & temperature, excessive
dust, or if the monitor has been abused in some way.
The procedure outlined below uses mono-plugs that contain an appropriate resistor for
testing the wrist strap input jacks located on the front of the 3M 732 & 733 Remote Input
Jacks. Use of this kit eliminates the need of a resistance substitution box when testing the
wrist strap input function. For additional information on the operation of the 724
Monitor see instruction manual.
Procedure:
Test Voltage 9 Volts & Resistance Selection 10 Megohms
Use plugs 1 through 4 on wrist strap inputs.
Set the 724 Test Voltage to 9 Volts and the Resistance Limit to 10 Megohms. In the order
outlined below insert each test plug into the #1 input of the Remote Input Jack and
observe the lamps and audible alarm on the 724 Monitor:
1. Test Plug #1 (1.33 Megohms) – Green (OK) lamp (1) ON, Yellow low lamp (L)
2. Test Plug #2 (1.69 Megohms) - Green (OK) lamp (1) ON, Yellow low lamp (L) OFF,
3. Test Plug #3 (8.45 Megohms) – Red high wrist strap lamp (H) OFF, Green (OK)
4. Test Plug #4 (11.5 Megohms) – Red high wrist strap lamp (H) ON, audible alarm ON
5. Repeat steps 1 through 4 for input #2 of the 732 Remote Input Jack. Audible alarm
flashing, and audible alarm OFF.
and audible alarm OFF.
lamp (1) ON, and audible alarm OFF.
(continuous tone), and Green (OK) lamp (1) OFF.
will be a fast chirping beep.
724VK Verification Kit
Verification Procedure for 3M™ Model 724 Workstation Monitor
Using the 3M Model 724 VK Verification Kit
1

724VK Summary of contents

Page 1

... Verification Kit Verification Procedure for 3M™ Model 724 Workstation Monitor Using the 3M Model 724 VK Verification Kit The 724 Workstation Monitor cannot be re-calibrated after the initial factory calibration. 3M recommends that the operation of the monitor be verified year interval or more frequently if used in harsh environments such as high humidity & temperature, excessive dust the monitor has been abused in some way ...

Page 2

Test Voltage 9 Volts & Resistance Selection 35 Megohms Use plugs and 6 on wrist strap inputs. Set the 724 Test Voltage to 9 Volts and the Resistance Limit to 35 Megohms. In the order outlined below ...

Page 3

... Test Plug #6 (40.2 Megohms) – Red high wrist strap lamp (H) ON, audible alarm ON (continuous tone), and Green (OK) lamp (1) OFF. 5. Repeat steps 1 through 4 for input #2 of the 732 Remote Input Jack. Audible alarm will be fast chirping beep. 3M Electronic Handling & Protection Division 6801 River Place Blvd. Austin, TX 78726-9000 www.3M.com/ehpd 3 ...