Page 1
Wrist strap + shoe tester Operating Instructions MO 0765-0601 3 740 wrist strap + shoe tester < > o. 0765-0601 April, 2001 ...
Page 2
LED for «too low resistance» Ground cord 4 mm stud Contact plate for wrist strap test Socket for wrist strap Ground Socket Output signal for external devices e.g. - data transfer to PC, electrical door opening system, acoustic signal, etc. ...
Page 3
... The 740 wrist strap and shoe tester consists of: - Instruction manual with daily log sheet - Base unit - Data output connector plug - Wall mounting kit, 3M Dual Lock™ fastening system and template - Cover for wrist strap plug-in jack - AC/DC transformer - Ground cord Accessories: ...
Page 4
... Make sure the area on the 740 in which the Dual Lock™ fastening strips are placed is free of dust and dirt. Attach the 3M Dual Lock™ fastening strips to the rear of the 740 tester to the bottom and top in parallel; see drawing for locations. ...
Page 5
VII. Shoe test The 741 shoe electrode has to be connected to the 740 shoe jack. Switch the 740 on. Stand on the 741 shoe electrode. If you are using shoe straps with 1 MΩ resistor on both feet to ...
Page 6
VIII. Data Output Connector If you are going to use the data output connector you must follow a test sequence. It must be wrist strap test first and then the shoe test. If this sequence is not strictly kept, the ...
Page 7
Data output connector pin status when testing wrist strap/shoe ...
Page 8
X. Verification Procedure The following procedure can be used to determine if the 740 operates within its specifications. Please note: The tester has no adjustable components. Equipment needed: 1. resistors 750 kΩ - 120 MΩ, tolerance ± ...
Page 9
Verification of shoe test circuit: Connect the reference resistor to the 740 as shown in Figure 2. Select the 10 MΩ range using the reference resistors indicated in the test table and press contact plate B. 740 wrist strap + ...
Page 10
... Wrist Strap + Shoe Tester Logbook Example: Test parameters: Wrist Strap (W) Not ok: 750 kΩ to: 2 MΩ Shoes (S) 750 kΩ to: 10 MΩ 5 MΩ 10 MΩ 35 MΩ Unit calibrated on: By: 35 MΩ Next calibration on: 50 MΩ Serial number: 100 MΩ ...
Page 11
... BUT NOT LIMITED TO, ANY IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. If this product is defective within the warranty period stated above, your exclusive remedy shall be, at 3M's option, to replace or repair the 3M product or refund the purchase price of the 3M product. Except where prohibited by law, 3M will not be liable for any loss or damage arising from this 3M product, whether direct, indirect, special, incidental or consequential regardless of the legal theory asserted ...