dtmpu30-101 Wall Industries,Inc., dtmpu30-101 Datasheet

no-image

dtmpu30-101

Manufacturer Part Number
dtmpu30-101
Description
30 Watts Single, Dual, And Triple Outputs Ac/dc Medical Desktop Power Supply
Manufacturer
Wall Industries,Inc.
Datasheet
FEATURES
● Splash Proof
● 3 Year Warranty
● Class I Insulation
● IEC-320-C14 Input Inlet
● Optional Output Connectors
● Single, Dual, and Triple Outputs
● Wide Input Voltage 90 to 264VAC, 47~63Hz
● Output Voltage Protection (Crowbar Design)
● Output Voltage Available from 3VDC thru 40VDC
● Input Surge Current, Over Voltage, and Over Load Protection
*Note: The Ripple & Noise for output voltages under 3.3VDC is 2% max.
SPECIFICATIONS: DTMPU30 Series
SPECIFICATION
INPUT (V
Operating Voltage Range
Input Frequency
Input Current (Low Line)
Input Current (High Line)
Inrush Current (Low Line)
Inrush Current (High Line)
Safety Ground Leakage Current
Start-Up Time
OUTPUT (V
Output Voltage Range
Load Regulation
Line Regulation
Output Power Range
Output Current Range
*Ripple & Noise (peak to peak)
Transient Response
Hold-Up Time
Temperature Coefficient
PROTECTION
Over Voltage Protection
Over Current Protection
GENERAL
Efficiency
Dielectric Withstanding Voltage
For Primary to Secondary
Dielectric Withstanding Voltage
For Primary to Ground
Isolation Resistance
ENVIRONMENTAL
Operating Temperature
Storage Temperature
Relative Humidity
PHYSICAL
Weight
Dimensions
Warranty
SAFETY
EMI Requirements for CISPR-11
EMI Requirements for FCC PART-18
in
)
All specifications are based on 25
o
)
We reserve the right to change specifications based on technological advances.
Wall Industries, Inc. 5 Watson Brook Road Exeter, NH 03833
603-778-2300 www.wallindustries.com Fax 603-778-9797
TEST CONDITIONS
Io = Full Load, Vin = 115VAC
Io = Full Load, Vin = 230VAC
Io = Full Load, 25°C, Cool Start, Vin = 115VAC
Io = Full Load, 25°C, Cool Start, Vin = 230VAC
Io = Full Load, Vin = 240VAC
Io = Full Load, Vin = 100VAC
Vin = 230VAC
Io = Full Load
Vin = 90 to 264VAC
Full Load, Vin = 90VAC
Io = Full Load to Half Load, Vin = 100VAC
Io = Full Load, Vin = 110VAC
All Outputs
Io = Full Load, Vin = 230VAC
Primary to Secondary
Primary to Ground
Test Voltage = 2100VDC
Derate linearly from 100% Load at 50°C to 50% load at 70°C
Vin = 220VAC
Vin = 110VAC
o
C, Nominal Input Voltage, and Maximum Output Current unless otherwise noted.
Rev. B
AC/DC Medical Desktop Power Supply
Single, Dual, and Triple Outputs
-0.04
5600
2800
Min
Approximately 400 ~ 460
112
110
-40
0.3
90
47
12
70
50
B
B
0
0
5
4.65 x 2.56 x 1.54
See Rating Chart
See Rating Chart
Nom
1.5
0.5
0.5
12
26
83
3
3
DTMPU30 Series
+0.04
Max
0.34
264
132
150
0.9
0.1
63
15
30
10
30
85
70
85
95
2
1
1
4
30 Watts
inches
grams
Unit
years
Class
Class
%/°C
VDC
VDC
VDC
VAC
mA
MΩ
Hz
ms
ms
°C
°C
W
%
%
%
%
%
%
%
A
A
A
A
A
s

Related parts for dtmpu30-101

dtmpu30-101 Summary of contents

Page 1

... Output Voltage Protection (Crowbar Design) ● Output Voltage Available from 3VDC thru 40VDC ● Input Surge Current, Over Voltage, and Over Load Protection SPECIFICATIONS: DTMPU30 Series All specifications are based reserve the right to change specifications based on technological advances. SPECIFICATION ...

Page 2

... OUTPUT VOLTAGE / CURRENT RATING CHARTS Model Number Preset Voltage DTMPU30-101 DTMPU30-102 DTMPU30-103 DTMPU30-104 11 VDC DTMPU30-105 13 VDC DTMPU30-106 16 VDC DTMPU30-107 21 VDC DTMPU30-108 27 VDC DTMPU30-109 33 VDC DTMPU30-110 40 VDC Model Number Output Voltage Output 1 +3.3 VDC DTMPU30-200 Output 2 +12.0 VDC Output 1 +5.0 VDC DTMPU30-201 Output 2 +12.0 VDC Output 1 +5 ...

Page 3

... DERATING CURVE MECHANICAL DRAWING Unit: inches [mm] Wall Industries, Inc. 5 Watson Brook Road Exeter, NH 03833 Rev. B 603-778-2300 www.wallindustries.com Fax 603-778-9797 DTMPU30 Series 30 Watts Single, Dual, and Triple Outputs AC/DC Medical Desktop Power Supply ...

Related keywords