365 TPI (Test Products Int), 365 Datasheet - Page 5
365
Manufacturer Part Number
365
Description
SINGLE INPUT LUMBERG K-TYPE
Manufacturer
TPI (Test Products Int)
Series
-r
Type
Handheldr
Specifications of 365
Temperature Range
-328 ~ 1112°F (-200 ~ 600°C)
Temperature Display
C°/F°
Display Type
LCD
Input Type
Thermocouple
Features
Auto Off
Probe Type
K
Battery Cell Size
AA (2)
Includes
-
Lead Free Status / Rohs Status
Lead free / RoHS Compliant
Other names
365TPI
Available stocks
Company
Part Number
Manufacturer
Quantity
Price
Part Number:
365
Manufacturer:
FAIRCHILD/仙童
Quantity:
20 000
Company:
Part Number:
36500-0002
Manufacturer:
MOLEX
Quantity:
31 752
Company:
Part Number:
36500-0008
Manufacturer:
MOLEX
Quantity:
31 753
Company:
Part Number:
36500-0026
Manufacturer:
MOLEX
Quantity:
31 754
Company:
Part Number:
36500-0027
Manufacturer:
MOLEX
Quantity:
31 757
Copyright © 2008 Test Products International, Inc.
Models 306C Penetration tip comes with A306
Pocket Digital Thermometers
Features and Specifications
FAQ
FEATURES
Water Resistant
Water Proof
Tip Type
Stem Length
Data Hold
°C/°F Switchable
Range
Accuracy
Resolution
IP Rating
Auto Off
Sample Time
Battery All TPI Digital Thermometers use an LR44 battery
"C" versions also available 306CX and 307CX; both units come with A306 protective rubber boot.
Pocket Digital Thermometers
Min. Temp°F
Min. Temp°C
Max. Temp°F
Max. Temp°C
°F
°C
How do I check calibration
of my pocket thermometer?
You put the thermometer in a
solution of crushed ice and
water, swirl the water around,
and it should read close to 32°F.
All “C” version digidials can be
Probe
Tips
307C Needle-tip comes with A306
306C* 307C*
penetration
Use for immersion and air;
however air response time will
be slower than if using an actu-
al air probe.
-40°F
300°F
0.1°F
1 sec
4.9
1%
NA
•
•
•
Penetration
needle
300°F
-40°F
0.1°F
1 sec
4.9
1%
NA
www.tpi-thevalueleader.com • 503.520.9197
A306
Protective rubber boot
for the 306C, 306CX,
307C, and 307CX.
•
•
•
penetration
312C
Does the whole stem need
to be immersed to get an
accurate reading?
The sensor is in the tip of the
probe and needs to be 1/2 inch
into what you are measuring.
0.1°F/C
300°F
150°C
-58°F
-50°C
±1°C
1 sec
±2°F
IP63
4.9
•
•
•
•
Need chisel tip for surface tem-
peratures. Penetration and/or air
tips will not give an accurate
surface reading.
penetration
314C
0.1°F/C
-50°C
300°F
150°C
-58°F
1 sec
±1°C
±2°F
IP67
4.9
•
•
•
•
Chisel
auto field calibrated
penetration
316C
0.1°F/C
150°C
-50°C
300°F
-58°F
1 sec
±2°F
±1°C
IP63
2.8”
•
•
•
•
Models 312C Penetration-Tip
How do I calibrate the “C” version digital thermometer?
1. Insert the stainless steel shaft of the thermometer into an
2. Press and hold the D-H/Cal button for approximately
0.1°F/C
318
300°F
150°C
chisel
-58°F
-50°C
±1°C
1 sec
±2°F
IP63
2.8”
ice water bath and allow the reading to stabilize
8 seconds until “CAL” is displayed. Calibration is complete.
•
•
•
•
Measures three times faster than
chisel tip.
314C Penetration-Tip
323
contact
0.1°F/C
319
-50°C
300°F
150°C
-58°F
1 sec
±1°C
±2°F
IP63
2.8”
Contact
•
•
•
•
Chisel-Tip
± 1°C (<0 and >70°C)
±2°F (<32 and >158)
± 0.5°C (0 to 70°C)
±1°F (32 to 158°F)
penetration
320C
0.1°F/C
150°C
-50°C
300°F
-58°F
1 sec
IP67
2.8”
in 32º F ice water to ± 2º F
Models 316C Penetration-Tip (Photo)
•
•
•
•
Use for immersion and semi-solids.
Low mass, quick response, small
diameter penetration
318
319
320C Small diameter (1.6mm)
0.1°F/C
323
300°F
150°C
chisel
-58°F
-50°C
1 sec
±1°C
±2°F
IP63
4.9”
•
•
•
•
326
329
Chisel-Tip
Contact-Tip
Penetration-Tip
Needle
0.1°F/C
Needle-Tip
Contact-Tip
326
needle
-50°C
300°F
150°C
-58°F
1 sec
±1°C
±2°F
IP63
4.9”
•
•
•
•
contact
0.1°F/C
329
150°C
-50°C
300°F
-58°F
1 sec
±2°F
±1°C
IP63
4.9”
•
•
•
•
5