GPM40B SL Power ( Ault / Condor ), GPM40B Datasheet - Page 2

no-image

GPM40B

Manufacturer Part Number
GPM40B
Description
Power Supply, Switching; 40 W; 5.1 V; 15 V; -15 V; 0 to degC; 1.0 lbs (Max.)
Manufacturer
SL Power ( Ault / Condor )
Type
Medicalr
Datasheet

Specifications of GPM40B

Brand/series
GPM40 Series
Configuration
Open-Frame
Mounting Type
Panel
Number Of Outputs
3
Operation
Switching
Output
-15 VDC @ 0.4 A, +5.1 VDC @ 4 A, +15 VDC @ 2 A
Power, Output
40 W
Power, Rating
40 W
Primary Type
AC-DC
Standards
UL Listed, CSA Certified and CE Marked
Temperature Coefficient
0.03 %⁄°C (Typ.)
Temperature, Operating
0 to +50 °C
Voltage, Input
85 to 264 VAC
Voltage, Output
-15, +5.1, +15 VDC
Voltage, Output 1
5.1 V
Voltage, Output 2
15 V
Voltage, Output 3
-15 V
Weight
1.0 lbs (Max.)
Note: For 48 volt model, GPC40-48 or GPM40-48, contact factory for availability.
A. Total regulation is defined as the maximum deviation from the nominal voltage for all steady-state conditions of initial voltage setting, input line voltage and output load.
B. To maintain these regulation conditions, the 5.1V current must be at least 1/4 of V2 and not greater than 5 times the V2 current.
C. Requires +5.1V to be adjusted within ±1% with at least a 0.4A load to maintain regulation on this output.
D. Add “G” suffix to model number for RoHS compliant model.
Commercial
Model
GPC40A
GPC40B
GPC40D
GPC40-3.3
GPC40-5
GPC40-9
GPC40-12
GPC40-15
GPC40-24
GPC40-28
ENVIRONMENTAL SPECIFICA-
TIONS
Temperature (A)
Humidity (A)
Shock (B)
Altitude
Vibration (C)
GPC40/GPM40 MECHANICAL SPECIFICATIONS
3
Medical
Model
GPM40A
GPM40B
GPM40D
GPM40-3.3
GPM40-5
GPM40-9
GPM40-12
GPM40-15
GPM40-24
GPM40-28
OPERATING
See individual specs.
0 to 95% RH
20 g
-500 to 10,000 ft
1.5 g
Output No.
pk
rms’
1
2
3
1
2
3
1
2
3
1
1
1
1
1
1
1
0.003 g
2
/Hz
Output
+5.1 V
+5.1 V
+5.1 V
+12 V
+15 V
+24 V
-12 V
-15 V
-12 V
3.3 V
12 V
15 V
24 V
28 V
5 V
9 V
NON-OPERATING
-40 to +85°C
0 to 95% RH
40 g
-500 to 40,000 ft
5 g
rms’
pk
0.026 g
2
/Hz
Minimum
Output
0.4 A
0.4 A
0.4 A
0 A
0 A
0 A
0 A
0.A
0.A
0 A
0 A
0 A
0 A
0 A
0
0
A. Units should be allowed to warm up/operate under non-condensing
B. Shock testing—half-sinusoidal, 10 ± 3 ms duration, ± direction,
C. Random vibration—10 to 2000Hz, 6dB/octave roll-off from 350 to
onditions before application of power.
3 orthogonal axes, total 6 shocks.
2000Hz, 3 orthogonal axes. Tested for 10 min./axis operating and
1 hr./axis non-operating.
Maximum
3
Output
0.4 A
0.4 A
0.4 A
4.5 A
3.3 A
2.7 A
1.7 A
1.4 A
4 A
2 A
4 A
2 A
4 A
1 A
8 A
8 A
Output
Peak
0.7 A
0.7 A
1.5 A
0.7 A
4.2 A
3.3 A
2.1 A
1.8 A
10 A
10 A
5 A
3 A
5 A
3 A
5 A
6 A
Noise P-P
120 mV
120 mV
150 mV
150 mV
240 mV
120 mV
120 mV
150 mV
240 mV
280 mV
50 mV
50 mV
50 mV
35 mV
50 mV
90mV
Regulation
Total
2%
5%
3%
2%
5%
3%
2%
5%
3%
2%
2%
2%
2%
2%
2%
2%
Notes
B,C
B,C
B,C
C
C
C

Related parts for GPM40B