N2735A OPTION 004 AGILENT TECHNOLOGIES, N2735A OPTION 004 Datasheet - Page 21
N2735A OPTION 004
Manufacturer Part Number
N2735A OPTION 004
Description
Oscilloscope Software
Manufacturer
AGILENT TECHNOLOGIES
Series
DSO7000r
Datasheet
1.N5406A.pdf
(28 pages)
Specifications of N2735A OPTION 004
No. Of Channels
4
Scope Bandwidth
350 MHz
Scope Type
Hand Held
Lead Free Status / RoHS Status
na
Performance characteristics
Trigger system (continued)
Nth edge burst
Autoscale
Scope channel triggering
Range (internal)
Sensitivity*
Coupling
Digital (D15 - D0) channel triggering (7000B Series MSO or MSO-upgraded 7000B Series DSO)
Threshold range (user defined)
Threshold accuracy
Predefined thresholds
External (EXT) triggering
Input impedance
Maximum input
Range
Sensitivity
Coupling
Probe ID
*
Denotes warranted specifications, all others are typical. Specifications are valid after a 30-minute warm-up period and ±10 °C from firmware calibration temperature.
Trigger on the Nth edge of a burst that occurs after an idle time that you specify. Max edge
count: 65,536.
Finds and displays all active scope and logic (for 7000B Series MSO) channels, sets edge
trigger mode on highest-numbered channel, sets vertical sensitivity on scope channels and
thresholds on logic channels, time base to display ~1.8 periods. Requires minimum voltage
>10 mVpp, 0.5% duty cycle and minimum frequency >50 Hz.
±6 div from center screen
<10 mV/div: greater of 1 div or 5 mV; ≥10 mV/div: 0.6 div
AC, ~10 Hz on MSO/DSO701xB/703xB/705xB/710xB,
DC, noise reject, HF reject and LF reject (~50 kHz)
±8.0 V in 10 mV increments
±(100 mV + 3% of threshold setting)
TTL = 1.4 V, CMOS = 2.5 V, ECL = –1.3 V
MSO/DSO7xx2B (2-/2+16-ch models)
1 MΩ ± 3% || 14 pF or 50 Ω
CAT I 300 Vrms, 400 Vpk, CAT II 100 Vrms, 400 Vpk
With 10073C/D 10:1 probe: CAT I 500 Vpk, CAT II 400 Vpk
5 Vrms with 50-Ω input
DC coupling: trigger level ±1 V and ±8 V
For ±1 V range setting:
DC to 100 MHz, 100 mV;
For ±8 V range setting:
DC to 100 MHz, 250 mV;
AC (~3.5 Hz), DC, noise reject, HF reject and LF reject (~50 kHz)
Auto probe sense and AutoProbe interface
Agilent- and Tektronix-compatible passive probe sense
>100 MHz to bandwidth of oscilloscope: 500 mV
(continued)
MSO/DSO7xx4B (4-/4+16-ch models)
±15 V
±5 V
DC to 500 MHz: 500 mV
2.14 kΩ ±5%
21